With interest growing in on-line microbial monitoring, Stilmas, a company specialized in the design, development, and construction of storage and distribution systems for pharmaceutical waters, collaborated with METTLER TOLEDO to assess our 7000RMS(TM) bioburden analyzer.
One of the objectives was to evaluate the quality of data from the analyzer as a process control measurement of microorganisms in routine water system operation. In addition, the 7000RMS was assessed to determine the efficiency and effectiveness of sanitization performed on the water system to maintain microbial control.
This case study looks at the results of Stilmas' testing of the 7000RMS and considers the analyzer's efficiency in microbial monitoring for the control of a pharmaceutical water system.
7000 RMS の仕組みを見てみましょう。細菌や真菌のような微生物には、成長と発育を調節するための重要な代謝物が含まれています。これらの代謝物は、特定の波長の光を受けると蛍光を発します。7000 RMS 微生物検出分析装置は、レーザー誘起蛍光を使用して、処理水中の微生物の存在を瞬時に検出します。 処理水は連続的に検査室へ引き込まれ、そこでレーザービームが照射されます。流れている水流中の粒子は光を散乱させます。散乱した光は放物線レンズと光検出器によって捕捉されます。代謝物を含む微生物細胞はレーザーによって励起され、蛍光として光エネルギーを放出します。 seven thousand rms 内の高度なアルゴリズムが、粒子信号と蛍光信号を組み合わせて微生物細胞を識別し、自家蛍光ユニット(afu)として分類します。afu のカウントは seven thousand rms 上に大きく表示され、連続的なリアルタイムデータとして記録されます。リアルタイムの微生物検出について詳しくは、www.mt .com/seventhousandrms をご覧ください。
The compendial test for microbial monitoring and control of a water system predominantly relies on the traditional plate-count method. This method not only requires a long time to obtain results, it may also introduce contamination to samples, possibly resulting in false-positive tests.
The 7000RMS bioburden analyzer is an at-line instrument that provides real-time and continuous microbial monitoring of a water system, without sample preparation or incubation. The units generated by the 7000RMS, auto-fluorescent units (AFUs), are a measurement of the microbial cells in a water system. As a sample of the water flows through the 7000RMS, it is exposed to a 405 nm wavelength laser. Metabolites, such as NaDH and riboflavin, on the membrane of any cells in the water emit fluorescence in response to the laser light. The analyzer detects the fluorescence and counts how many cells are present. This approach eliminates the variability and time-consuming nature of traditional plate counting. Real-time microbial monitoring allows water release without having to wait for plate-count results.


.png/_jcr_content/renditions/cq5dam.web.1280.1280.png)
.png/_jcr_content/renditions/cq5dam.web.1280.1280.png)

.png/_jcr_content/renditions/cq5dam.web.1280.1280.png)