The USP outlines three stages of testing conductivity for bulk waters:

Stage 1

On-line temperature and non-temperature-compensated conductivity are measured. The measured temperature value is rounded down to the next lowest 5°C interval. If the measured conductivity is not greater than the limit outlined in Table 1 of USP<645>, the water meets the requirements of USP <645>.

Stage 2

This stage requires taking a sample and stirring it to equilibrate it to atmospheric CO2. If the conductivity is less than 2.1 μS/cm at 25°C ± 1°C, it passes; if not, move to Stage 3. 

Stage 3

Saturate the sample from Stage 2 with potassium chloride (KCl). Measure pH to the nearest 0.1 pH unit. Look up the conductivity limit for that pH in Table 2 in USP <645> or in the table below. If the measured conductivity (from Stage 2) is less than the conductivity limit, it passes. If the measured conductivity exceeds this value or the pH is outside the range of 5.0–7.0, the water does not meet the requirements of USP <645> for conductivity.

In a conductivity measurement, there are three main components that are responsible for measuring or affecting the conductivity:

Temperature

The first is the temperature. Conductivity is highly temperature-dependent, so conductivity readings are useless without knowing the water or liquid’s temperature. It is worth noting that using an external temperature sensor is an acceptable method for collecting temperature data, so if a conductivity sensor has an embedded temperature sensor it does not need to be calibrated in this scenario. However, in most cases, an embedded temperature sensor is used to avoid additional equipment.

Cell Constant

The second part is the cell constant. This is a multiplier applied to the measurement based on the sensor's electrode geometry. A sensor with a greater volume of water between the electrodes will have lower conductivity, since the electrical current must travel through more liquid. The cell constant allows the sensor to report the conductivity of the water per centimeter or other distance units. This allows two different conductivity sensors with different cell constants to report conductivity values in the same units, allowing readings to be compared.

Measurement Electronics

The third part is the measurement electronics. These are responsible for measuring the resistance between the two electrodes. This resistance is then converted into conductivity, as they are inversely proportional. The measurement electronics can be embedded in the sensor's transmitter for analog sensors or embedded directly in the sensor for digital sensors. The pharmaceutical industry historically used analog sensors; however, many companies are switching to digital sensors for increased accuracy and stability.

Temperature must be accurate to ± 2°C

Cell constant is accurate and known to ± 2%

Then either:

  • Calibrate the sensor in a solution with stated conductivity (from NIST, chemical supplier, etc.). or
  • Calibrate the sensor in a solution prepared to a specific conductivity (ASTM D1125 standard or ultrapure water). or
  • Calibrate sensor vs. another calibrated sensor, usually from the same manufacturer.
Bốn trụ cột của chất lượng nước dược phẩm

Nắm vững việc tuân thủ USP <645>

Cách Giám Sát Độ Dẫn Điện Theo Thời Gian Thực Đảm Bảo Độ Tinh Khiết Của Nước Sử Dụng Trong Ngành Dược

Bốn trụ cột của chất lượng nước dược phẩm

USP <643> Total Carbon hữu cơ: Cách tiếp cận thông minh hơn để tuân thủ

Giám sát tổng cacbon hữu cơ (TOC) theo thời gian thực hợp lý hóa việc kiểm soát chất lượng nước cho sản xuất dược phẩm như thế nào

Giải thích về cảm biến độ dẫn điện

Giải thích về cảm biến độ dẫn điện

Từ cảm biến hai điện cực đến cảm biến cảm ứng: Hiểu các loại cảm biến độ dẫn điện, cách thức hoạt động và ứng dụng của chúng

Tôi muốn…
Bạn cần hỗ trợ?
Đội ngũ của Chúng tôi luôn sẵn sàng.