SmartCheck
Material No.: 30448961
Material No.: 30448967
Verify any Brand of Pipette
Use SmartCheck to verify the performance of any brand of pipette that dispenses 10 –1,000 μL. Also verify individual channels of multichannel pipettes.
Small Footprint
SmartCheck fits easily on any bench. The compact size enables a pipette to be tested to ensure it performs within tolerance limits before starting an experiment.
Avoid Rework
Verifying a pipette before working saves precious reagents and time. You know if a pipette is aspirating and dispensing correctly within 60 seconds
Material No.: 30564095
Product Type | Pipette Verification Device |
Repeatability typical (sd) | 0.08 μL @ 100 μL |
Vessel Capacity | 10 mL |
Measurement Principle | Gravimetric, 4 volumetric measures |
Pipette Test Volume | 10 µL 20 µL 50 µL 100 µL 200 µL 300 µL 1,000 µL |
Pipette Test Tolerance | Default ± 5%, adjustable 3 -10% |
Repeatability Limit (sd) | 0.15 μL @ 1'000 μL |
Reduce Risk
In regulated environments, discovering a pipette is out of tolerance can lead to a time-consuming and expensive investigation. SmartCheck provides reassurance that the pipette you are using is performing as expected.
Trust Your Pipette
Between calibrations, SmartCheck provides a fast, practical way to verify that a pipette is dispensing accurately. Simply dispense distilled water into the opening and the test volume will be automatically detected. Repeat this three times to obtain a pass/fail result
Material No.: 30448961
Material No.: 30448967