Sodium Measurement Range
| 0.001 - 100,000 ppb or equivalent ppm, autoranging |
Sodium Measurement Accuracy
| ±10% of reading or ± 0.05 ppb whichever is greater
|
Response Time
| 90% at 5 minutes |
Sample pH
| 2.5-12 |
Update Rate
| Once per second |
Reagent Consumption
| Diisopropylamine (DIPA), approximately 1 L per 3 months more at higher temperatures and for cation exchange samples
|
Sample Flow rate
| >40 mL/min (>20 mL/min for cation exchange samples), excess to drain |
Sample Temperature
| 5-50°C (41-122°F) |
Sample Pressure
| 0.3-7 bar (5-100 psig) |
Grab Sample Measurement
| Included |
Power Supply
| 100-240 VAC, 50-60 Hz, 25 W, on power loss all settings are retained without batteries |
Ambient Operating Temperature
| 10-45°C (50-113°F) |
Humidity
| 10-90% non-condensing |
Process Connection
| Drain Hose: 19x25.4 mm (3/4 x 1") 2 m (6 ft) length included Sample Inlet: 1/4" or 6 mm OD tube SS compressin fitting Drain Hose: 19x25.4 mm (3/4 x 1") 2 m (6 ft) length included
|
Operator Interface
| 4-line backlit LCD, 5 tactile keys simultaneous display of sodium, conditioned pH, auto-cal status (temperature optional)
|
Analog Output
| 22mA alarm, 500 ohm max load, not for use with externally powered circuit For sodium, conditioned pH, temperature four powered 0/4-20 mA,
|
Analog Output Scaling
| Linear, bi-linear, logarithmic (1,2,3,4 decades), auto ranging |
Measurement Parameter
| Sodium |
Electrode Conditioning
| Part of auto-cal sequence |
Relay Contacts
| 3A resistive freely assignable to setpoints for sodium, pH, temperature other relays used for auto-cal
|
Sodium Calibration
| Automatic, unattended 3-point known addition |
Calibration
| Automatic |
Analog Output Accuracy
| ±0.05 mA |
Model Family
| 2300Na |
EnclosureType
| Panel |
Built-in sequencer
| No |