Cond Sensor 3/4NPT 0.1C Ti/SS 2

Cond Sensor 3/4NPT 0.1C Ti/SS 2

2-Electrode Analog Conductivity Sensor.This analog conductivity sensor is a 2-electrode sensor. Due to its titanium concentric electrodes, this sensor is suitable for measurement in ultra-pure and high-purity water.

Water Measurement in Multiple Industries

The 2-electrode sensor is suitable for pharmaceutical water, power and steam generation, semiconductor water treatment and recycle-and-reclaim water.

Traceable Calibration

Sensor accuracy is assured with ASTM and NIST traceable calibration and certification of the cell constant and temperature sensor.

Very Fast Response

The titanium electrode of this conductivity sensor makes it more sensitive, so it responds more quickly to measurements.

Material No.: 58031201

Call for Quote

Features

Exceptional Measurement Accuracy

The optimized 4-wire measuring circuitry provides exceptional range and accuracy, eliminating cable effects.

Strong Chemical Resistance

Chemical resistance and an optimal response are provided through titanium electrodes, the PTFE-coated stainless steel fitting and the PEEK insulator.

For Most Pure and Ultrapure Water Types

The NPTM fitting on the conductivity sensor is designed for most standard conductivity applications. This makes it suitable for the vast majority of pure and ultrapure water applications.

Specifications

Specifications - Cond Sensor 3/4NPT 0.1C Ti/SS 2

Operating Range (µS/cm)
0.02 – 2,000
Temperature Sensor
PT 1000
Cell Constant
0.1 cm-1
Connection
¾ inchNPT
Cable
1.5 ft (0.5 m)
Connector
Standard
Measurement Technology
Conductive
Fitting material
PTFE/SS
Insertion length
34mm
Model Family
Analog
Hazardous Area Approvals
No
Measured Pressure Resistance
250 psig @ 200 °F (17 bar @ 93 °C)
Electrode Material
Titanium
Primary Media/Application
Other Industrial Processes
Other Water Processes
Power Plant Water/Steam
Semiconductor Water
Hygienic/Biocompatibility Approvals
No
Process fitting
¾" NPTM
Cell Constant (accuracy)
± 1%
Electrode Configuration
2-electrode
Measurement Range
0.02 - 2,000 µS/cm
Sensor Material
Titanium, 316L SS

Documentation

Brochures

POWERFUL SOLUTIONS for Optimizing Your Power Plant Chemistry (NEW EDITION!)
The new edition of our Power Industry brochure addresses all areas of power plants where precise water analytics are needed. Get an overview of METTLE...

Datasheets

Data Sheet: Thornton ISM Sensors for M800 and M300 ISM
Digital liquid analysis sensors optimized for pure water treatment.

Accessories

Consumables

Software

More Information

Get Your Quote
Cond Sensor 3/4NPT 0.1C Ti/SS 2
FDA