Application Note
Requirements for in-line humidity validation, parallel to a standard routine measurement:
- free sample dish on the sample carousel,
- constant 25 °C measurement temperature,
- sorption and desorption cycle, ideally over an extended RH range, e.g. from 0 % RH to 90 % RH,
- addition of an extra 3 % RH step to the sorption cycle
During sample preparation, 100 mg of the MCC reference are distributed evenly on the bottom of the sample pan designated for the humidity validation. The other sample and measurement settings are prepared and started according to the applying standard operation procedure (SOP).