Hi, my name is Holly Nury, and today I'd like to introduce you to the new XPR105 MCP from Mettler Toledo. The MCP is Mettler Toledo's new high-speed calibration device for multichannel pipettes. With a readability of 10 micrograms and a repeatability of 15 micrograms, This system completely fulfills the ISO 8655 requirements for multichannel pipettes with nominal volumes from 0.5 to 1000 microliters. The supplementary Calibri PC software guides you step by step through pipette calibration procedures in order to ensure accurate and traceable calibrations according to ISO 8655, JJJ 646, and 21 CFR part 11. Now, let's take a look at how easy it is to complete a calibration with the MCP and Calibri. In the Calibri software, I can navigate to see my pipettes and then choose the one I will calibrate from the list. There are pre-configured tests for this pipette volume, or I can define a custom one. In this case, I chose the ISO 8655 calibration and can now follow along with the step-by-step guidance on the screen. As I pipette, Calibri automatically performs the calculations required for calibration. The status bar along the top tracks my progress, and it indicates with a red or green color whether the measurements are currently performing within specifications. Once your calibration is complete, a report is automatically generated and stored within Calibri. If you'd like, you can also create custom reports or export your reports to be saved externally. The system has one single load cell to ensure fast and equal performance across all pipette channels. It can evaluate the results of a single dispense in under one minute, meaning you can perform the 360 measurements required to calibrate a 12-channel pipette in as little as 25 minutes. The tanks can be easily emptied with either a pump or by manually emptying the tanks into a waste container. Additionally, two MCPs can be connected for increased productivity, enabling you to complete two simultaneous measurements. With this configuration, it takes as little as 15 minutes to complete the 30 dispenses required for a full calibration. In conclusion, the MCP is the perfect option for calibrations and performance verifications of multichannel pipettes. It fulfills all of the requirements of ISO 8655 #METTLERTOLEDO #MultichannelPipetteCalibration #MultichannelPipettes
The Gravimetric Reference Method
The gravimetric reference measurement procedure for the accurate determination of volume is described in part 6 of ISO 8655:2022.
The balance used for testing shall be chosen according to the minimum requirements specified in ISO 8655-6:2022, Table 1, depending on the nominal volume of the apparatus under test. The balance parameters are defined so that the expanded uncertainty in use is less than one-fourth of the maximum permissible systematic error of the apparatus.