With interest growing in on-line microbial monitoring, Stilmas, a company specialized in the design, development, and construction of storage and distribution systems for pharmaceutical waters, collaborated with METTLER TOLEDO to assess our 7000RMS(TM) bioburden analyzer.
One of the objectives was to evaluate the quality of data from the analyzer as a process control measurement of microorganisms in routine water system operation. In addition, the 7000RMS was assessed to determine the efficiency and effectiveness of sanitization performed on the water system to maintain microbial control.
This case study looks at the results of Stilmas' testing of the 7000RMS and considers the analyzer's efficiency in microbial monitoring for the control of a pharmaceutical water system.
Vediamo come funziona il 7000 RMS. Microrganismi come batteri e funghi contengono metaboliti chiave che regolano la crescita e lo sviluppo. Questi metaboliti fluorescono quando sono esposti a luce di specifiche lunghezze d'onda. L'analizzatore per il rilevamento microbico 7000 RMS utilizza la fluorescenza indotta dal laser per rilevare istantaneamente la presenza microbica nell'acqua di processo. L'acqua di processo viene continuamente convogliata in una camera di analisi, dove viene illuminata da un fascio laser. Le particelle presenti nel flusso d'acqua provocano la dispersione della luce. La luce diffusa viene captata da una lente parabolica e da un fotodetettore. Le cellule microbiche contenenti metaboliti vengono eccitate dal laser, rilasciando energia luminosa sotto forma di fluorescenza. Algoritmi sofisticati all'interno del seven thousand rms combinano i segnali delle particelle e della fluorescenza per identificare le cellule microbiche, classificandole come unità di autofluorescenza (afu). Il conteggio delle afu viene visualizzato in modo ben evidente sul seven thousand rms e registrato come dati continui in tempo reale. Per saperne di più sul rilevamento microbico in tempo reale, visita www.mt .com/seventhousandrms.
The compendial test for microbial monitoring and control of a water system predominantly relies on the traditional plate-count method. This method not only requires a long time to obtain results, it may also introduce contamination to samples, possibly resulting in false-positive tests.
The 7000RMS bioburden analyzer is an at-line instrument that provides real-time and continuous microbial monitoring of a water system, without sample preparation or incubation. The units generated by the 7000RMS, auto-fluorescent units (AFUs), are a measurement of the microbial cells in a water system. As a sample of the water flows through the 7000RMS, it is exposed to a 405 nm wavelength laser. Metabolites, such as NaDH and riboflavin, on the membrane of any cells in the water emit fluorescence in response to the laser light. The analyzer detects the fluorescence and counts how many cells are present. This approach eliminates the variability and time-consuming nature of traditional plate counting. Real-time microbial monitoring allows water release without having to wait for plate-count results.