The water content of active pharmaceutical ingredients (APIs) is a key parameter in quality control, helping to ensure shelf life and proper function when used in medical treatment. KF titration is the method of choice for these water content determinations because it is selective for water.
To ensure accurate determination of water content, it is important that samples possess good solubility in KF solvents and release water completely. There are various methods to address poor solubility, including the introduction of solubilizers to the titration cell. However, this approach has limitations, including the potential for side reactions with the sample, and the risk of decreased cell conductivity if solubilizers are added beyond a certain percentage. This challenge can be overcome by introducing heat to the sample solution.
In this article, we will discuss a Karl Fischer (KF) titration method that can improve the solubility of sample solutions, leading to more precise and efficient analysis. Specifically, we will examine a direct addition technique that utilizes a thermostatable vessel. By adopting this technique, we can achieve better solubility, resulting in more complete water release and more accurate determination of water content in antibiotics.
Experimental
In order to increase the solubility of the Amoxicillin aliquot, the EVA V3 titrator and a special thermo vessel (thermostatable vessel, as shown in Figure 1) were used to perform the measurement. The temperature was set to 50 °C. After an initial heating period (~20 minutes), ~50 mg powdered sample was added directly into the titration cell with a weighing boat and the KF titration started. A mixing time of 120 seconds ensured dissolution, and the correct sample weight was obtained by back-weighing. This was repeated six times.
![]() |
Figure 1. EVA V3 Karl Fischer Titrator and a thermostatable vessel. |
Results and discussion
In order to dissolve the sample solution, a thermostatable vessel was utilized. However, applying heat has the potential to cause over-titration. To prevent this potential issue, the EVA titrator with the cautious titrant addition was used. By implementing these parameters, accurate titration results can be achieved.
The results of water content analysis for all six antibiotic aliquots using direct injection volumetric titration are presented in Table 1. It indicates a mean water content of 9.55 ± 0.25%.
According to the results displayed in Figure 2, the KF titration reactions of the six samples reached their endpoint within a range of 60–65 seconds. These results suggest that this technique enables fast and efficient analysis.
![]() |
Table 1. Results of water content for six Amoxycillin aliquots. |
![]() |
Figure 2. Titration curves of six Amoxycillin aliquots (current in µA vs Time in seconds). |
Conclusion
The METTLER TOLEDO EVA V3 volumetric Karl Fischer titrator is a highly effective solution for determining water content in challenging samples such as those that are insoluble. With its thermostable KF vessel and advanced features, this system delivers fast and reliable results with outstanding precision and accuracy.
Pozdravljeni vsi. Danes vam želim predstaviti aplikacijo za titracijo Calfisher z našimi titratorji EVA. Gre za določanje vode v antibiotikih. Antibiotiki so razmeroma zahteven vzorec zaradi svoje omejene topnosti v običajnih topilih za reagente Calfisher, kot so posušijo nos. Zato tukaj uporabljamo segrevanje, da povečamo topnost in izboljšamo ekstrakcijo vode. Ta termostatska posoda je povezana s termostatom, ki bo krožil grelni medij skozi to termalno posodo, in običajno se tukaj uporablja 50 stopinj Celzija; segrevanje reaktiva traja približno 20 minut, zato je treba teh 20 minut počakati, potem pa je mogoče dodati antibiotike. Ena težava, ki se lahko pojavi pri segrevanju reagenta Calfisher, je tako imenovana prekomerna titracija zaradi povečane mobilnosti joda in povečane prevodnosti v raztopini vzorca. Zato imamo za to namenske kontrolne parametre, ki jih lahko izberete kot previdne. Ti previdni parametri omogočajo popolno titracijo pri določanju vode brez prekomerne titracije. Zdaj bi rad začel metodo in dodal približno 50 miligramov antibiotika v prahu. Zato zdaj tukaj začenjam metodo. In tukaj bom dodal vzorec antibiotika, približno 50 miligramov. Zdaj bom tehtalno posodico vrnil na tehtnico in masa vzorca bo samodejno prenesena tukaj v uporabniški vmesnik, vi pa jo lahko preprosto potrdite. Titracija se bo zdaj začela z mešalnim časom 60 sekund. Ta mešalni čas je pomemben, da se antibiotiki raztopijo, pa tudi da se vzorčni matriksu omogoči dovolj časa za sprostitev vode. Ta mešalni čas zdaj teče. Tukaj vidite števec, ki odšteva od 60 sekund do nič. Sekund. Prav tako vidite, da je izmerjena vrednost padla z 30 mikroamperov, kar je končna točka te amperometrične Koffischerjeve titracije, na nič, ker je bil ves jod porabljen, in zdaj mora titrator pri titraciji ponovno doseči končno točko 30 mikroamperov, to pa bo izvedeno previdno, da bi se preprečila kakršna koli prekomerna titracija. To zdaj poteka. Tukaj lahko vidite, da se poraba titranta začne. Drift je zelo visok. To je zato, ker se zdaj v titracijsko celico volumetrično dodaja veliko joda in doslej izmerjena vsebnost vode je že nad 2,7 miligrama, dosega tri miligrame, tok pa se tukaj počasi povečuje. Vidite, da se v celico zdaj ponovno dodaja vedno več joda, in takoj ko tukaj dosežemo 30 mikroamperov, se bo titracija ustavila. Kot dodaten kriterij prekinitve imamo relativno ustavitev zaradi drifta. Relativna ustavitev zaradi drifta pomeni, da drifta prištejete začetnemu driftu, ki ga določite v kriterijih prekinitve, in nato se bo titracija prav tako ustavila. Pri titraciji vedno začnete v suhih pogojih. Za to določite določen dovoljen začetni drift. Ta se lahko razlikuje glede na vrsto aplikacije. Za nekatere aplikacije lahko začnete z višjim začetnim driftom, za druge z nižjim začetnim driftom. Pomembno je, da drift ostane konstanten. To je najpomembnejši del. Zdaj tukaj vidimo, da je že doseženih 25 mikroamperov in kmalu bomo tukaj dosegli končno točko 30 mikroamperov. Vsebnost vode je nad 5,6 miligrama in zdaj pričakujem, da se bo titracija kmalu zaključila. Tukaj vidite, da drift pada. In zdaj imamo tukaj končni rezultat, ki bi moral za te antibiotike biti med 9 in 11 %. Torej smo na višjem koncu, vendar je to povsem razumen rezultat. Ta previdna titracija je uspela. In tako imamo tukaj lepo rešitev z uporabo termalne posode in namenskih titracijskih parametrov za to vrsto vzorca. Najlepša hvala in se vidimo kmalu.