Medium pressure crucible with pin
The medium-pressure crucible with pin can only be used for DSC measurements, the pan can also be used without lid.
Single use crucible
When sealed with a lid and special seal, it is a single use crucible and cannot be regenerated. It withstands a maximum pressure of 2 MPa.
Perfect crucible for liquids
The medium-pressure crucible is the perfect tool for samples containing liquids such as water, for temperatures above boiling point.
رقم المادة: 26929
Volume | 120 µL |
Maximum Temperature | 250 °C |
Number of Pieces | 25 |
Beaker Volume | 120 µL |
Special die/plunger set required
In order to seal the crucible, the crucible sealing press must be equipped with a special die set (ArtNo: 119428).
Seal from FPM is included
The medium pressure crucible is sealed with a FPM O-ring. FPM is slightly permeable to water and, therefore, this seal is more suitable for other liquids.
PCTFE seals for water
If water causes an issue, it is recommended to use PCTFE (polytrifluorochloroethylene) seals, Art. No. 26933. PCTFE, however, shows a DSC melting peak at about 220 °C.
Crucible with self-generated atmosphere
The crucible can also be used without a seal (sealed with lid only). In this case, a self-generated internal atmosphere will forms.