Crucible HP 40 uL, SST/Au, 25 pcs

HP 40 uL steel crucibles.Set of 25 crucibles (without pin), lids and rupture disks. The gold-plated steel crucibles have proven highly effective in the field of safety investigations. They can only be used for one measurement each. The maximum pressure is 15 MPa.

High-pressure crucibles up to 15 MPa

The high-pressure 40 μL crucibles can withstand an internal pressure of 15 MPa. This makes them perfect for use with highly reactive materials.

Gold-plated steel crucibles

The steel crucibles (exact material: X2 CrNiMo18143) are plated with a 5 µm gold coat. This avoids undesired reactions between sample and crucible.

Single-use crucible

Sealing the lids makes these crucibles so tight that they cannot be reopened again. Therefore, they cannot be reused.

رقم المادة: 26731

طلب أسعار
المواصفات - Crucible HP 40 uL, SST/Au, 25 pcs
Volume
40 µL
Maximum Temperature
400 °C
Number of Pieces
25
Beaker Volume
40 µL

HP crucibles with a perfectly flat base

The flat base ensures perfect thermal contact and is therefore suitable for enthalpy determinations.

Sealing mechanism

The crucible lid is pressed onto the crucible with the aid of a toggle press at a pressure of around one ton so that the rupture disk (Gold, Au 700/531) hermetically seals the crucible.

Tool kit / toggle press

A special tool kit (Art. No. 26733) for use with a toggle press. The Maeder toggle press (KP2.1N) is recommended.

HP crucible for robotic use

This type of high-pressure crucible is suitable for robotic use.