Now let's have a closer look on what's new. So what does this mean for you? The updates bring clear guidance on calibration, minimum weight, repeatability and accuracy. And they introduce a risk-based lifecycle approach to balance management. Here are the key points. Calibration is mandatory and must include measurement uncertainty. Calibration is required before and after any operation that could affect balance performance. Periodic checks on sensitivity and repeatability are now required between calibrations with frequency determined on a risk-based approach. And for the first time, a concrete safety factor is introduced. Two for stable lab conditions with trained operators, 1.5 for automated weighing procedures. In short, It's no longer just about passing a test. It's about proving your balance is consistently fit for purpose. How does Mettler Toledo support you? At Mettler Toledo, we have helped shape these principles through GWP, Good Weighing Practice, a scientific framework that supports the entire life cycle of your weighing instruments. Our experts are ready to guide you through the new USP requirements. They will provide the right certificates, Assist with risk assessment and help define optimal calibration intervals and routine testing frequencies. So you stay compliant without extra effort or unnecessary cost. Visit our website to download the white paper, watch the webinar and connect with our experts. Stay compliant, stay confident. With Mettler Toledo, you're ready for the new USP chapters 41 and 1251.
The revised mandatory Chapter 41 includes the following requirements:
- Calibration must be performed periodically, including measurement uncertainty, with frequency based on risk.
- Calibration is required before and after any operation that may affect balance performance (as found and as left).
- Periodic risk-based performance checks on sensitivity and repeatability must be conducted between calibrations.
- EURAMET cg-18 and ASTM E898 are mentioned as applicable standards for balance calibration