Sodium Measurement Range
| 0.01-100,000 ppb or equivalent ppm, auto-ranging |
Sodium Measurement Accuracy
| ± 10% of reading ± 0.1 ppb, typical using DIPA as reagent ± 10% of reading ± 1 ppb, typical using ammonium hydroxide as reagent
|
Response Time
| 90% at 5 minutes |
Sample pH
| 2.5-12 |
Update Rate
| Once per second |
Reagent Consumption
| 3-4 L per year DIPA, or 9-12 L per year ammonium hydroxide typical more at higher temperatures and for cation exchange samples
|
Sample Flow rate
| >40 mL/min (>20 mL/min for cation exchange samples), excess to drain |
Sample Temperature
| 5-50°C (41-122°F) |
Sample Pressure
| 0.3-7 bar (5-100 psig) |
Grab Sample Measurement
| Available option |
Power Supply
| 100-240 VAC, 50-60 Hz, 25 W, on power loss all settings are retained without batteries |
Dimensions (HxWxD)
| 851 x 400 x 165 mm (33.5" x 15.75" x 6.5") |
Weight
| 4.5 kg (10 lbs) |
Ambient Operating Temperature
| 5-50 °C (41-122 °F) |
Humidity
| 10-90% non-condensing |
Ratings/Approvals
| CE, UL, IP65 |
Dimensions (HxWxD)
| 851 mm x 400 mm x 165 mm |
Process Connection
| Drain Hose: 19x25.4 mm (3/4 x 1") 2 m (6 ft) length included Sample inlet: 6 mm or 1/4" OD tube SS compression fitting Drain hose: 19 x 25.4 mm (¾ x 1"), 2 m (6 ft) length included
|
Operator Interface
| 4-line backlit LCD, 5 tactile keys simultaneous display of sodium, conditioned pH, auto-cal status (temperature optional)
|
Analog Output
| 22mA alarm, 500 ohm max load, not for use with externally powered circuit For sodium, conditioned pH, temperature four powered 0/4-20 mA,
|
Power Supply (AC/DC)
| 100 V – 240 V |
Analog Output Scaling
| Linear, bi-linear, logarithmic (1,2,3,4 decades), auto ranging |
Flow Rate
| 40 mL/min – 60 mL/min |
Measurement Parameter
| Sodium |
Electrode Conditioning
| Manual |
Operating Range (Temp.)
| 5 °C – 50 °C |
Relay Contacts
| 3A resistive freely assignable to setpoints for sodium, pH, temperature other relays used for auto-cal
|
Sodium Calibration
| Manual 3-point known addition manual 1- or 2-point
|
Operating Range
| 0.01 ppb – 100,000 ppb |
Analog Output Accuracy
| ± 0.05 mA |
EnclosureType
| Full |
Outputs
| 4 Analog |
Built-in sequencer
| No |