To estimate the stability of such products and formulations, the knowledge about the amorphous content and the recrystallization kinetics is crucial.
In literature, several methods have been described to determine the amorphous content of powdered products. However, the detection limits of these methods are often limited to amorphous contents above 10 % for X-ray diffraction and Differential Scanning Calorimetry or 1 % for FT-Raman spectroscopy, NIR-spectroscopy, and solution calorimetry. Lower detection limits could be achieved using isothermal microcalorimetry (0.2 %) and solid-state NMR (0.5 %) [8].
Besides these methods gravimetric dynamic water vapor sorption (DVS) was successfully applied for determining amorphous contents. Compared to the above-mentioned methods, DVS is an easy and sensitive method with a detection limit down to 0.1 %.