Comprehensive SOPs and Acceptance Criteria

This guide supports the Easy pH, Easy Cl, Easy Ox, Easy Pro, and Easy KFV models:

  • Acid-Base (Easy pH/Pro): Standardize NaOH with KHP; the mean titer must fall between 0.975–1.025 with RSD ≤ 1.0%.
  • Argentometric (Easy Cl/Pro): Focuses on NaCl and silvernitrate using the EM45-BNC silver ring sensor.
  • Redox (Easy Ox/Pro): Standardizing sodium thiosulphate; using the "Cautious" control mode is recommended for better repeatability.
  • Karl Fischer (Easy KFV): Covers drift stability (< 25 μg/min) and water content recovery limits of 97.5–102.5%.
  • Technical Execution: Always rinse the burette twice and tap tubing to remove air bubbles before starting.
  • Traceable Reporting: Configure method parameters to "Export: PDF report" and "Printout: Long" for permanent records.