The water content of active pharmaceutical ingredients (APIs) is a key parameter in quality control, helping to ensure shelf life and proper function when used in medical treatment. KF titration is the method of choice for these water content determinations because it is selective for water.
To ensure accurate determination of water content, it is important that samples possess good solubility in KF solvents and release water completely. There are various methods to address poor solubility, including the introduction of solubilizers to the titration cell. However, this approach has limitations, including the potential for side reactions with the sample, and the risk of decreased cell conductivity if solubilizers are added beyond a certain percentage. This challenge can be overcome by introducing heat to the sample solution.
In this article, we will discuss a Karl Fischer (KF) titration method that can improve the solubility of sample solutions, leading to more precise and efficient analysis. Specifically, we will examine a direct addition technique that utilizes a thermostatable vessel. By adopting this technique, we can achieve better solubility, resulting in more complete water release and more accurate determination of water content in antibiotics.
Experimental
In order to increase the solubility of the Amoxicillin aliquot, the EVA V3 titrator and a special thermo vessel (thermostatable vessel, as shown in Figure 1) were used to perform the measurement. The temperature was set to 50 °C. After an initial heating period (~20 minutes), ~50 mg powdered sample was added directly into the titration cell with a weighing boat and the KF titration started. A mixing time of 120 seconds ensured dissolution, and the correct sample weight was obtained by back-weighing. This was repeated six times.
![]() |
Figure 1. EVA V3 Karl Fischer Titrator and a thermostatable vessel. |
Results and discussion
In order to dissolve the sample solution, a thermostatable vessel was utilized. However, applying heat has the potential to cause over-titration. To prevent this potential issue, the EVA titrator with the cautious titrant addition was used. By implementing these parameters, accurate titration results can be achieved.
The results of water content analysis for all six antibiotic aliquots using direct injection volumetric titration are presented in Table 1. It indicates a mean water content of 9.55 ± 0.25%.
According to the results displayed in Figure 2, the KF titration reactions of the six samples reached their endpoint within a range of 60–65 seconds. These results suggest that this technique enables fast and efficient analysis.
![]() |
Table 1. Results of water content for six Amoxycillin aliquots. |
![]() |
Figure 2. Titration curves of six Amoxycillin aliquots (current in µA vs Time in seconds). |
Conclusion
The METTLER TOLEDO EVA V3 volumetric Karl Fischer titrator is a highly effective solution for determining water content in challenging samples such as those that are insoluble. With its thermostable KF vessel and advanced features, this system delivers fast and reliable results with outstanding precision and accuracy.
歡迎各位。今天我想向大家展示使用我們 EVA 滴定儀進行的卡爾費休滴定應用。這是抗生素中的水分測定。抗生素因為在傳統卡爾費休試劑溶劑中的溶解度有限,所以屬於相對棘手的樣品,例如 會使鼻子乾燥。這就是為什麼我們在這裡採用加熱,以提高溶解度並增加水分萃取。這個恆溫容器連接到一個熱循環系統,該系統會將加熱介質循環通過這個熱容器,通常這裡會使用 50 攝氏度,加熱到 試劑大約需要 20 分鐘,因此這 20 分鐘需要等待,之後才能加入抗生素。加熱卡爾費休試劑時可能出現的一個問題是所謂的過滴定,這是由於碘的流動性提高以及樣品溶液導電度增加所致。因此,針對這一點,我們提供了專門的控制參數,您可以選擇為謹慎模式。 這些謹慎參數可在不過滴定的情況下,實現完美的水分滴定測定。所以現在我想開始方法,加入大約 50 毫克的抗生素粉末。現在我在這裡啟動方法。 然後我會在這裡加入抗生素樣品,約 50 毫克。現在我會把稱量舟放回天平上,樣品重量將自動傳輸到這裡的 GUI,您現在只需直接接受即可。 接著滴定將以 60 秒的混合時間開始。這段混合時間很重要,目的是使抗生素溶解,同時也給水分從樣品基質中釋放出來足夠的時間。現在混合時間正在進行中。您可以看到這裡有一個計數器從 60 秒倒數到零。 秒。您也可以看到量測值已從 30 微安培下降到零;30 微安培是這種電量法卡爾費休滴定的終點,因為所有碘都已被消耗。現在在滴定過程中,滴定儀必須再次達到 30 微安培的終點,而這將以謹慎的方式進行,以便 避免任何過滴定。現在流程正在進行中。您可以看到這裡開始消耗滴定劑。我們有非常高的漂移。這是因為此時大量碘以體積方式加入到滴定池中,而目前量得的含水量已超過 2.7 毫克,達到 3 毫克,且電流正在這裡緩慢上升。 您會看到越來越多的碘又被加入到池中,一旦我們在這裡達到 30 微安培,滴定就會停止。我們還有一個額外的終止條件,即相對漂移停止。相對漂移停止是指將漂移加到您在終止條件中定義的起始漂移上,然後滴定也會停止。 進行滴定時,您總是從乾燥狀態開始。為此您要定義一個允許的起始漂移。這可能會因應用類型而有所不同。因此,對某些應用您可以使用較高的起始漂移開始,對其他應用則使用較低的起始漂移。重要的是漂移保持不變。這是最重要的部分。現在我們在這裡看到 已經達到 25 微安培,很快就會達到 30 微安培的終點。含水量已超過 5.6 毫克,我預期滴定很快就會完成。請看這裡漂移正在下降。 現在我們在這裡得到了最終結果,這些抗生素的結果應介於 9% 和 11% 之間。所以我們落在較高的一端,但這是完全合理的結果。這次謹慎滴定已完成。因此,我們在這裡得到了一個很好的解決方案,使用熱容器和針對這種樣品類型的專用滴定參數。非常感謝,我們下次再見。