Operating Range (Temp.)
| 0 °C – 37 °C |
Operating Range (Humidity)
| 10 % – 90 % |
Flow Rate
| 30 mL/min – 100 mL/min |
Dimensions (HxWxD)
| 616 mm x 564 mm x 305 mm |
Analysis Time
| Continuous |
Wall Mount
| Anti-vibration shelf required (P/N 58 079 700) |
Minimum detection size
| ≥ 0.3 μm |
Inlet pressure
| 21 - 80 psig (2 - 5.5 bar(g)) Calibration, cleaning and grab sample requires sample pressure of 0 psig (0 bar(g) Process pressure above 80 psig (5.5 bar(g)) requires optional High Pressure Regulator (P/N 58 091 552)
|
Humidity (non-condensing)
| 80% maximum relative humidity up to 31 °C (87.8 °F) decreasing linearly to 50% relative humidity at 40 °C (104 °F) |
Ambient temperature (non-condensing)
| 1 - 37 °C (32 - 98.6 °F) Temperature below 15 °C or above 45 °C requires Sample Conditioning Coil (included)
|
Voltage
| MAINS supply voltage fluctuations up to ±10% of the nominal voltage of 100 - 240 VAC 50 - 60 Hz TEMPORARY OVERVOLTAGES occurring on the MAINS SUPPLY TRANSIENT OVERVOLTAGES: up to levels of OVERVOLTAGE CATEGORY II
|
Data communication
| 5 - 20 mA standard, with configurable output ranges Analog output channels Ethernet - standard RJ 45 / Wi-Fi capable SCADA connectivity via Modbus TCP USB
|
Enclosure Material
| Stainless Steel |
Measurement Range
| 1 -10,000 AFU / mL |
Response Time
| 3 seconds (1 mL) |
Physical dimensions
| (WxHxD) 22.2"(56.4 cm) W x 24.25"(61.6 cm) H x 12"(30.5 cm) D |
Monitoring location
| At-line to drain |
Flow rate
| 31 mL/min |
Use
| Indoor use |
Environmental Temperature
| 6 - 35 °C (41- 95 °F) |
Weight
| 73.4 lbs (33.3 kg) |
Outlet connection
| 0.125" (3 mm) O.D. |
Inlet connection
| 0.125" (3 mm) O.D. |
Power Requirements
| 100 - 240 VAC 50 - 60 Hz 5A Use the power cord included with the instrument 8.2' (2.5 m) cord length provided standard
|
Temperature (non-condensing)
| 6 - 90 °C (41-194 °F) Temperature below 15 °C or above 45 °C requires Sample Conditioning Coil (included) |
Environment
| Pollution degree 3 |
Detection limit
| 2 AFU (Auto Fluorescent Units) |
Type / Quality
| Purified Water (PW), Ultrapure Water (UPW), Water for Injection (WFI) |
Altitude
| Up to 6562' (2000 m) |