Single Use pH Sensor | Single Use Sensor | Disposable pH Sensor
 
메뉴
Single-use Sensors

Single-use Sensors for Bag Manufacturers

pH and Dissolved Oxygen for stirred Single-use Reactors (> 20 L)

Designed for Single-Use Bag Manufacturers
Easy and Safe Installation and Integration
Reliable, Straightforward Use in Biocontrollers

제품 & 기술사양

 
제품 & 기술사양
필터:
검색 조건 설정
Clear All
 
Short description
측정범위
자세히 보기
Short descriptionConventional pH sensor with pH sensitive membrane glass
측정범위pH 3-10
감지부Glass pH and Reference Electrodes, Lead-Free
통합형 온도 전극Pt1000
영점7.20 ±0.25 pH
검교정1-point process
샤프트 재질High density polyethylene, USP 88 class VI
O-ring 재질EPDM, USP 88 class VI
Process connectionEldon James 1' port disc, with retainer
감마선 멸균 가능25…45 kGy
Operating Temperature5…60°C (41…140°F)
Storage temperature5…25°C (41…77°F)
기계적 허용 압력"2 barg / 40 °C (29 psig / 104 °F), 1 barg / 60 °C (15 psig / 140 °F)"
승인USP 88 class VI for all wetted polymer parts, factory calibration
자세히 보기
Short descriptionOptical dissolved oxygen sensor
측정범위0…250% air
감지부Optical spot (Fluorescensce quenching)
통합형 온도 전극Pt1000
검교정1-point air
샤프트 재질High density polyethylene, USP 88 class VI
O-ring 재질EPDM, USP 88 class VI
Process connectionEldon James 1' port disc
감마선 멸균 가능25…45 kGy
Operating Temperature5…60°C (41…140°F)
Storage temperature5…25°C (41…77°F)
기계적 허용 압력"2 barg / 40 °C (29 psig / 104 °F), 1 barg / 60 °C (15 psig / 140 °F)"
승인USP 88 class VI for all wetted polymer parts
비교하기

문서

Application Note

Application Note: Differences in Operation Between Analytical Multi-Use and Single-Use Sensors in Bioreactors
This application note provides information on the important trend towards the deployment of single-use bioreactors (SUBs) in the bioprocess industry....

기술 사양

Technical Data: InSUS x07 Sensor Series
The InSUS™ x07 sensor series for single-use technology bioprocesses cover the measurement parameters pH and dissolved oxygen (DO).

Instruction Manuals

Instruction Manual: InSUS Head for Optical Single-Use O₂ Sensors
These operating instructions contain all the information needed for safe and proper use of the optical sensor head.
Instruction Manual: Information for InSUS fiber optic cable
The enclosed adapter must be used, if the fiber optic cable is connected to sensors of the type InSUS 607. Mount the adapter – as shown in the photos...
M300 / M400 2G Quick Calibration Guide For pH Single-Use Sensors Type InSUS 307
METTLER TOLEDO InSUS™ pH sensors are analog, gamma irradiation-sterilizable single-use pH sensors with an integrated Pt 1000 temperature probe.
Thank you for visiting www.mt.com. We have tried to optimize your experience while on the site, but we noticed that you are using an older version of a web browser. We would like to let you know that some features on the site may not be available or may not work as nicely as they would on a newer browser version. If you would like to take full advantage of the site, please update your web browser to help improve your experience while browsing www.mt.com.