The water content of active pharmaceutical ingredients (APIs) is a key parameter in quality control, helping to ensure shelf life and proper function when used in medical treatment. KF titration is the method of choice for these water content determinations because it is selective for water.
To ensure accurate determination of water content, it is important that samples possess good solubility in KF solvents and release water completely. There are various methods to address poor solubility, including the introduction of solubilizers to the titration cell. However, this approach has limitations, including the potential for side reactions with the sample, and the risk of decreased cell conductivity if solubilizers are added beyond a certain percentage. This challenge can be overcome by introducing heat to the sample solution.
In this article, we will discuss a Karl Fischer (KF) titration method that can improve the solubility of sample solutions, leading to more precise and efficient analysis. Specifically, we will examine a direct addition technique that utilizes a thermostatable vessel. By adopting this technique, we can achieve better solubility, resulting in more complete water release and more accurate determination of water content in antibiotics.
Experimental
In order to increase the solubility of the Amoxicillin aliquot, the EVA V3 titrator and a special thermo vessel (thermostatable vessel, as shown in Figure 1) were used to perform the measurement. The temperature was set to 50 °C. After an initial heating period (~20 minutes), ~50 mg powdered sample was added directly into the titration cell with a weighing boat and the KF titration started. A mixing time of 120 seconds ensured dissolution, and the correct sample weight was obtained by back-weighing. This was repeated six times.
![]() |
Figure 1. EVA V3 Karl Fischer Titrator and a thermostatable vessel. |
Results and discussion
In order to dissolve the sample solution, a thermostatable vessel was utilized. However, applying heat has the potential to cause over-titration. To prevent this potential issue, the EVA titrator with the cautious titrant addition was used. By implementing these parameters, accurate titration results can be achieved.
The results of water content analysis for all six antibiotic aliquots using direct injection volumetric titration are presented in Table 1. It indicates a mean water content of 9.55 ± 0.25%.
According to the results displayed in Figure 2, the KF titration reactions of the six samples reached their endpoint within a range of 60–65 seconds. These results suggest that this technique enables fast and efficient analysis.
![]() |
Table 1. Results of water content for six Amoxycillin aliquots. |
![]() |
Figure 2. Titration curves of six Amoxycillin aliquots (current in µA vs Time in seconds). |
Conclusion
The METTLER TOLEDO EVA V3 volumetric Karl Fischer titrator is a highly effective solution for determining water content in challenging samples such as those that are insoluble. With its thermostable KF vessel and advanced features, this system delivers fast and reliable results with outstanding precision and accuracy.
皆さま、ようこそ。本日は、当社のEVA滴定装置を用いたCalfisher滴定のアプリケーションをご紹介したいと思います。これは抗生物質中の水分測定です。抗生物質は、通常のCalfisher試薬用溶媒では溶解度が限られているため、比較的扱いの難しい試料です。たとえば 鼻を乾かすようなものです。そのため、ここでは溶解度を高め、水の抽出を増やすために加熱を行っています。この恒温容器はサーモスタットに接続されており、加熱媒体がこの恒温容器内を循環します。通常はここで50℃を使用し、加熱には約20分かかります。 試薬を加温しますので、この20分間は待ってから抗生物質を添加します。Calfisher試薬を加熱した際に起こりうる問題の一つが、ヨウ素の移動度の増加と試料溶液の導電率上昇に起因する、いわゆる過滴定です。そのため、慎重モードとして選択できる専用の制御パラメータを用意しています。 これらの慎重パラメータにより、過滴定のない完全な滴定による水分測定が可能になります。では、ここで方法を開始し、粉末状の抗生物質を約50ミリグラム添加したいと思います。これから方法を開始します。 そしてここに抗生物質試料を約50ミリグラム添加します。次に、秤量皿を天びんに戻すと、試料質量が自動的にここGUIへ転送され、あとは簡単に承認できます。 そして、滴定は60秒の攪拌時間で開始されます。この攪拌時間は、抗生物質を溶解させるためだけでなく、試料マトリクスから水を十分に遊離させるためにも重要です。現在、この攪拌時間が進行しています。60秒から0までカウントダウンするカウンターが見えます。 秒。測定値も、電量滴定であるこのCalfisher滴定の終点である30マイクロアンペアから、ヨウ素がすべて消費されたためゼロまで下がっているのが分かります。これから滴定では、滴定装置が再び30マイクロアンペアの終点に到達する必要があり、それは過滴定を避けるために慎重に行われます。 これは現在進行中です。ここで滴定液の消費が始まっているのが分かります。非常に大きなドリフトがあります。これは、多量のヨウ素がここで滴定セルに容量的に添加され、これまでに測定された水分量がすでに2.7ミリグラムを超え、3ミリグラムに達しつつあり、電流が徐々に増加しているためです。 ヨウ素がさらに多くセルに追加されており、ここで30マイクロアンペアに到達すると、滴定は停止します。追加の終了判定基準として、相対ドリフト停止があります。相対ドリフト停止とは、終了条件で定義した開始ドリフトに現在のドリフトを加算し、その値に達すると滴定も停止するという意味です。 滴定を行うときは、常に乾燥状態から開始します。そのため、ある許容開始ドリフトを設定します。これはアプリケーションの種類によって異なります。あるアプリケーションでは開始ドリフトを高めに、別のアプリケーションでは低めに設定して開始できます。 重要なのは、ドリフトが一定に保たれることです。それが最も重要な点です。ここで、 すでに25マイクロアンペアに達しており、まもなく30マイクロアンペアの終点に到達します。 水分量は5.6ミリグラムを超えており、まもなく滴定が完了すると見込んでいます。 ここでドリフトが下がってきているのが分かります。 そして今、最終結果が得られました。これらの抗生物質では9〜11%の範囲であるべきです。したがって、やや高めではありますが、十分に妥当な結果です。この慎重な滴定が達成されました。 このように、サーモ容器とこの試料タイプ向けの専用滴定パラメータを用いた、優れたソリューションを得ることができました。ご清聴ありがとうございました。またお会いしましょう。