Analyzer 2300Na Sodium Full Door - Overview - METTLER TOLEDO

Analyzer 2300Na Sodium Full Door

Early Detection of Trace Sodium.The 2300Na Sodium Analyzer provides early detection of trace sodium contamination in a water system with minimal operator intervention. It is an ideal solution for in-line sodium monitoring in ultrapure water or steam cycle chemistry.

Accurate, Reliable Sodium Monitoring

Combination sodium and pH electrode system confirms reagent addition for reliable operation, minimizes reagent consumption, and reduces maintenance.

Configurable, Automated Calibration

Unattended, automated calibration ensures the accurate measurement of sodium in your water system and detects breakthroughs right when they occur.

Fully Enclosed for Dirty Environments

Especially suitable for plant floor locations in coal-fired power plants and where a locking door is needed.

Material No.: 58042002

Call for Quote
Specifications - Analyzer 2300Na Sodium Full Door
Sodium Measurement Range
0.001 - 100,000 ppb or equivalent ppm, autoranging
Sodium Measurement Accuracy
±10% of reading or ± 0.05 ppb
whichever is greater
Response Time
90% at 5 minutes
Sample pH
2.5-12
Update Rate
Once per second
Reagent Consumption
Diisopropylamine (DIPA), approximately 1 L per 3 months
more at higher temperatures and for cation exchange samples
Sample Flow rate
>40 mL/min (>20 mL/min for cation exchange samples), excess to drain
Sample Temperature
5-50°C (41-122°F)
Sample Pressure
0.3-7 bar (5-100 psig)
Grab Sample Measurement
Included
Power Supply
100-240 VAC, 50-60 Hz, 25 W, on power loss all settings are retained without batteries
Dimensions (HxWxD)
35.4"x17.7"x7.5" (900x450x190 mm)
Weight
27 kg (60 lbs)
Ambient Operating Temperature
10-45°C (50-113°F)
Humidity
10-90% non-condensing
Process Connection
Sample Inlet: 1/4" or 6 mm OD tube SS compressin fitting
Drain Hose: 19x25.4 mm (3/4 x 1") 2 m (6 ft) length included
Operator Interface
4-line backlit LCD, 5 tactile keys
simultaneous display of sodium, conditioned pH, auto-cal status (temperature optional)
Analog Output
For sodium, conditioned pH, temperature
four powered 0/4-20 mA,
22mA alarm, 500 ohm max load, not for use with externally powered circuit
Analog Output Scaling
Linear, bi-linear, logarithmic (1,2,3,4 decades), auto ranging
Measurement Parameter
Sodium
Electrode Conditioning
Part of auto-cal sequence
Relay Contacts
3A resistive freely assignable to setpoints for sodium, pH, temperature
other relays used for auto-cal
Two unpowered, SPDT, 250VAC/30VDC
Sodium Calibration
Automatic, unattended 3-point known addition
Calibration
Automatic
Analog Output Accuracy
±0.05 mA
Model Family
2300Na
EnclosureType
Full
Built-in sequencer
No

Sub-PPB Sodium Monitoring at a Glance

The 2300Na Sodium Analyzer delivers more than monitoring at sub-ppb sodium concentration levels. The simultaneous display of temperature, adjusted pH and calibration progress provide a complete picture to increase uptime and reduce operating costs.

Greater Accuracy (Unattended)

The 2300Na Sodium Analyzer eliminates the need for operator intervention with pre-configured, periodic self-calibration. Providing greater measurement accuracy, an internal pH measurement ensures interference-free sodium determination.

Full Protection in Dirty Environments

The fully enclosed 2300Na Sodium Analyzer protects all components, preventing damage from outside interference. This ensures safe, reliable and continuous operation in production environments.

Low Maintenance Made Easy

With 6-month preventive maintenance intervals, the 2300Na Sodium Analyzer requires less downtime and reduces the overall critical operator time commitment. Every calibration automatically results in sensor conditioning, further reducing maintenance requirements.

Documentation

Datasheets

2300Na Sodium Analyzer
The 2300Na Sodium Analyzer delivers more than monitoring at sub-ppb sodium concentration levels. The simultaneous display of temperature, adjusted pH...

Accessories

Consumables

Software

Get Your Quote
Analyzer 2300Na Sodium Full Door
FDA
USP Class VI