Cond Sensor 2Tri 4E SS PEEK VP

Cond Sensor 2Tri 4E SS PEEK VP

4-Electrode Analog Conductivity Sensor.This analog conductivity sensor is a 4-electrode sensor. Due to its highly polished 316L stainless steel electrodes, this sensor is suitable for hygienic applications such as pharmaceutical water.

Water Measurement in Multiple Industries

The 4-electrode sensor is suitable for pharmaceutical water, recycle-and-reclaim water and wastewater treatment.

Traceable Calibration

Sensor accuracy is assured with ASTM and NIST traceable calibration and certification of the cell constant and temperature sensor.

Suitable for Hygienic Applications

The highly polished 316L stainless steel electrode and tri-clamp fitting make this conductivity sensor especially suitable for hygienic applications.

Material No.: 58031243

Call for Quote

Features

Exceptional Measurement Accuracy

The optimized 4-wire measuring circuitry provides exceptional range and accuracy, eliminating cable effects.

Strong Chemical Resistance

Chemical resistance and an optimal response are provided thanks to the 316L stainless steel electrodes and PEEK body.

Fitting Meets Hygienic Requirements

The conductivity sensor's tri-clamp fitting makes it suitable for applications with hygienic requirements.

Specifications

Specifications - Cond Sensor 2Tri 4E SS PEEK VP

Operating Range (µS/cm)
10 – 650,000
Temperature Sensor
PT 1000
Cell Constant
4-E
Connection
VP (IP 68) , 2 inch Tri-Clamp
Connector
VarioPin
Measurement Technology
Conductive
Fitting material
PEEK
Insertion length
25mm
Model Family
Analog
Hazardous Area Approvals
No
Measured Pressure Resistance
70 psig @ 302 °F and 200 psig @ 122 °F (5 bar @ 150 °C and 14 bar @ 50 °C)
Electrode Material
SS 316L
Primary Media/Application
Other Industrial Processes
Other Water Processes
Pharmaceutical Waters
Pharmaceutical/Life Science Production
Wastewater
Hygienic/Biocompatibility Approvals
Yes
Process fitting
2" Tri-Clamp
Cell Constant (accuracy)
± 5% system accuracy
Electrode Configuration
4-electrode
Measurement Range
10 - 650,000 µS/cm
Sensor Material
PEEK/SS

Documentation

Brochures

POWERFUL SOLUTIONS for Optimizing Your Power Plant Chemistry (NEW EDITION!)
The new edition of our Power Industry brochure addresses all areas of power plants where precise water analytics are needed. Get an overview of METTLE...

Datasheets

Data Sheet: Thornton ISM Sensors for M800 and M300 ISM
Digital liquid analysis sensors optimized for pure water treatment.

Accessories

Consumables

Software

More Information

Get Your Quote
Cond Sensor 2Tri 4E SS PEEK VP
FDA