How to Measure pH in Small Samples
Know How
Application Method
Tips and Tricks

How to Measure pH in Small Samples

Know How
Application Method
Tips and Tricks

Various applications in life sciences involve very small samples. With the right pH micro sensor it is possible to minimize sample size and sample manipulation to achieve accurate pH measurement results.

pH Measurement Basics

Various applications in life sciences involve very small samples that require analysis by multiple protocols like pH measurement. These samples typically represent costly experiments, and sample manipulation can introduce error in the results. With the right pH micro sensor it is possible to minimize sample size and manipulation to achieve accurate results.

Some small-volume applications:

  • RNA sample analysis using microarrays
  • DNA sample analysis using PCR
  • Protein sample analysis using immunoprecipitation

Table of content:

  1. The challenges of pH measurement in small samples
  2. How to measure pH in small samples
  3. pH micro sensors / pH surface sensors
Get accurate pH results in your small samples

1. The challenges of pH measurement in small samples

There are three key components influencing the accuracy of the pH measurement: the sensing glass, the reference junction and the temperature input.
The sensing glass is the membrane at the bottom of a pH sensor. It is doped with ions allowing it to generate a voltage potential when it interacts with hydronium ions in solution. The resulting voltage potential is then converted into a useable pH reading. In order to generate an accurate voltage potential, the sensing glass must be in full contact with the solution being analyzed.
Partial exposure of the membrane to air during measurement will introduce error into the result. For this reason, standard sized sensors are not appropriate for use with very small samples. The sample does not completely cover the sensing glass, leaving an air/ glass interface.
The reference junction is typically located just above the sensing glass. This allows reference electrolyte to flow out of the sensor and into the sample. Without steady electrolyte flow into the sample, error in the reading is unavoidable. Small samples rarely meet the electrolyte junction of a standard sized sensor, which prevents the flow of electrolyte into the sample and yields erroneous results.
pH results are only correct if the sample temperature is taken into account during the measurement. In very small samples the sample mass is negligible compared with the sensor mass. Thus, the sensor can take so long to reach equilibrium that the sensor temperature is wrongly interpreted as the sample temperature.

2. How to measure pH in small samples

Because the pH value of a sample represents the concentration of hydronium in the solution, it is frequently impossible to dilute the sample and still yield an accurate pH value. In such cases, the pH value must be obtained from the original sample solution, and specialized sensors can be employed to generate accurate data.

Figure 1: Standard sized sensors pictured next to equivalent micro and surface sensors.
Figure 1: Standard sized sensors pictured next to equivalent micro and surface sensors.

3. pH micro sensors / pH surface sensors

 
 
 
 
 
 
 
Thank you for visiting www.mt.com. We have tried to optimize your experience while on the site, but we noticed that you are using an older version of a web browser. We would like to let you know that some features on the site may not be available or may not work as nicely as they would on a newer browser version. If you would like to take full advantage of the site, please update your web browser to help improve your experience while browsing www.mt.com.