The water content of active pharmaceutical ingredients (APIs) is a key parameter in quality control, helping to ensure shelf life and proper function when used in medical treatment. KF titration is the method of choice for these water content determinations because it is selective for water.
To ensure accurate determination of water content, it is important that samples possess good solubility in KF solvents and release water completely. There are various methods to address poor solubility, including the introduction of solubilizers to the titration cell. However, this approach has limitations, including the potential for side reactions with the sample, and the risk of decreased cell conductivity if solubilizers are added beyond a certain percentage. This challenge can be overcome by introducing heat to the sample solution.
In this article, we will discuss a Karl Fischer (KF) titration method that can improve the solubility of sample solutions, leading to more precise and efficient analysis. Specifically, we will examine a direct addition technique that utilizes a thermostatable vessel. By adopting this technique, we can achieve better solubility, resulting in more complete water release and more accurate determination of water content in antibiotics.
Experimental
In order to increase the solubility of the Amoxicillin aliquot, the EVA V3 titrator and a special thermo vessel (thermostatable vessel, as shown in Figure 1) were used to perform the measurement. The temperature was set to 50 °C. After an initial heating period (~20 minutes), ~50 mg powdered sample was added directly into the titration cell with a weighing boat and the KF titration started. A mixing time of 120 seconds ensured dissolution, and the correct sample weight was obtained by back-weighing. This was repeated six times.
![]() |
Figure 1. EVA V3 Karl Fischer Titrator and a thermostatable vessel. |
Results and discussion
In order to dissolve the sample solution, a thermostatable vessel was utilized. However, applying heat has the potential to cause over-titration. To prevent this potential issue, the EVA titrator with the cautious titrant addition was used. By implementing these parameters, accurate titration results can be achieved.
The results of water content analysis for all six antibiotic aliquots using direct injection volumetric titration are presented in Table 1. It indicates a mean water content of 9.55 ± 0.25%.
According to the results displayed in Figure 2, the KF titration reactions of the six samples reached their endpoint within a range of 60–65 seconds. These results suggest that this technique enables fast and efficient analysis.
![]() |
Table 1. Results of water content for six Amoxycillin aliquots. |
![]() |
Figure 2. Titration curves of six Amoxycillin aliquots (current in µA vs Time in seconds). |
Conclusion
The METTLER TOLEDO EVA V3 volumetric Karl Fischer titrator is a highly effective solution for determining water content in challenging samples such as those that are insoluble. With its thermostable KF vessel and advanced features, this system delivers fast and reliable results with outstanding precision and accuracy.
欢迎大家。今天我想向大家展示一个使用我们的 EVA 滴定仪进行的 Calfisher 滴定应用。它用于抗生素中的水分测定。抗生素是一类相对难处理的样品,因为它们在常规 Calfisher 试剂溶剂中的溶解度有限,例如 会使鼻腔干燥。因此我们在这里采用加热,以提高溶解度并增强水分提取。这个恒温容器连接到一个恒温系统,该系统会让加热介质在这个恒温容器中循环。通常这里使用 50 摄氏度,加热大约需要 20 分钟 以加热 试剂,因此需要等待这 20 分钟,之后才可以加入抗生素。加热 Calfisher 试剂时可能出现的一个问题是所谓的过度滴定,这是由于碘的流动性增加以及样品溶液电导率升高所致。因此,我们为此提供了专门的控制参数,您可以选择为谨慎模式。 这些谨慎参数可实现完美的滴定水分测定,而不会过度滴定。现在我想启动该方法,并加入大约 50 毫克的粉末状抗生素。所以我现在在这里启动该方法。 然后我将在这里加入抗生素样品,大约 50 毫克。现在我会把称量舟放回天平上,样品重量会自动传输到这里的 GUI 中,您现在只需直接确认即可。 随后滴定将以 60 秒的混合时间开始。这个混合时间很重要,目的是使抗生素充分溶解,同时也给予足够时间使样品基质中的水分释放出来。现在混合时间正在进行。您可以在这里看到一个计数器从 60 秒倒数到零。 秒。您还可以看到测量值已经从 30 微安下降到零,30 微安是这种安培法卡尔费休滴定的终点,因为所有碘都已被消耗;而现在在滴定过程中,滴定仪必须再次达到 30 微安的终点,并且会以谨慎的方式进行,以便 避免任何过度滴定。现在滴定正在进行。您可以看到这里开始消耗滴定剂。漂移非常高,这是因为此时有大量碘以体积方式加入到滴定池中,而截至目前测得的含水量已经超过 2.7 毫克,达到 3 毫克,并且电流在这里缓慢上升。 您可以看到越来越多的碘再次加入到池中,一旦这里达到 30 微安,滴定就会停止。作为额外的终止标准,我们还有相对漂移停止。相对漂移停止是指将漂移加到您在终止条件中定义的起始漂移上,然后滴定也会停止。 进行滴定时,您总是从干燥条件开始。为此您需要定义一个允许的起始漂移。该值可能因应用类型而异。因此,对于某些应用,您可以使用较高的起始漂移开始,而对于其他应用则使用较低的起始漂移。重要的是漂移保持恒定。这是最重要的部分。现在我们看到这里 已经达到 25 微安,很快这里就会达到 30 微安的终点。含水量已超过 5.6 毫克,我预计滴定很快就会完成。请看这里,漂移正在下降。 现在我们这里得到了最终结果,对于这些抗生素,结果应在 9% 到 11% 之间。因此我们处于较高一端,但这是完全合理的结果。这次谨慎滴定取得了成功。因此,我们在这里得到了一个很好的解决方案,使用恒温容器以及针对这种样品类型的专用滴定参数。非常感谢,期待很快再见。