Chloride Measurement Range
| 0-300 ppb |
Chloride Limit of Detection
| 0.5 ppb |
Chloride Measurement Accuracy
| ±5% of reading or ± 0.5 ppb whichever is greater
|
Sulfate Measurement Range
| 0-300 ppb |
Sulfate Limit of Detection
| 1 ppb |
Sulfate Measurement Accuracy
| ±10% of reading or ±1 ppb whichever is greater
|
Measurement Cycle Time
| 45 min typical, programmable between 15 minutes and 1 hour |
Sample Flow rate
| 25-50 mL/min |
Sample Temperature
| 10-45°C (50-113°F) |
Sample Pressure
| 0.3-7 bar (5-100 psig) |
Grab Sample Measurement
| Included 100 mL capacity
|
Power Supply
| 100-240 VAC, 50-60 Hz, 100 W typical |
Dimensions (HxWxD)
| 927x508x305 mm (36.5"x20.9"x12") |
Weight
| 44 kg (97 lbs) |
Ambient Operating Temperature
| 10-35°C (50-95°F) |
Humidity
| 10-70% non-condensing |
Ratings/Approvals
| CE / cULus |
Process Connection
| Sample Inlet: 1/4" or 6 mm OD tube SS compressin fitting Drain Hose: 19x25.4 mm (3/4 x 1") 2 m (6 ft) length included
|
Operator Interface
| Color touchscreen Simultaneous display of ion concentrations and analyzer status
|
Analog Output
| 22mA alarm, 500 ohm max load, not for use with externally powered circuit |
Analog Output Scaling
| Linear, bi-linear, logarithmic (1,2,3,4 decades), auto ranging |
Measurement Parameter
| Chloride and sulfate |
Relay Contacts
| Mechanical rated at 250 VAC, 3 Amps (Relay 1 NC, Relay 2 to 4 NO), 4-SPDT Type Reed 250 VAC or DC, 0.5 Amps (Relay 5 to 8) |
Analog Output Accuracy
| ±0.05 mA |
Model Family
| 3000CS |
EnclosureType
| Full |
Built-in sequencer
| No |