Measurement Parameter
| Oxygen |
Measurement Technology
| Amperometric/Polarographic |
Accuracy
| ± (1 % of the reading + 6 ppb) |
Response Time
| T98 = < 90 s @ 25°C (77°F)(air to N2) |
Primary Media/Application
| Pharmaceutical/Life Science Production |
IP Rating
| IP68 |
ISM (Digital) or Analog
| ISM (Digital) |
Anti-Bubble Design
| No |
Wetted Metallic Material
| 316L stainless steel |
Surface Roughnes of Wetted Material
| N5/Ra16 (Ra=0.4μm/16μin) |
Sensor Diameter
| 12mm |
O-Ring Material
| Silicone |
Process Connection
| Pg 13.5 |
Sensor Length
| 70 mm |
Wetted membrane material
| PTFE |
Connector Design
| Straight |
Model Family
| InPro 6800/6850i |
Autoclavable/Sterilizable
| Yes |
CIP resistant
| Yes |
Mechanical Temp. Range
| 23 °F – 284 °F (-5 °C – 140 °C) |
Operating Pressure (Bar)
| 0.2–6 Bar |
Operating Pressure (PSI)
| 0–87 PSI |
Design Pressure
| 174 psi (12 bar) |
Cable Connector
| K8S |
Hygienic/Biocompatibility Approvals
| EHEDG/FDA/USP Class 7 |
Hazardous Area Approvals
| ATEX:/IECEx: Ex ia IIC T6/T5/T4/T3 Ga/Gb/Ex ia IIIC T69°C/T81°C/T109°C/T161°C Da/Db FM (C US): IS / I, II, III / 1 / ABCDEFG/ T6 Ta=60 °C 53 800 002 Entity
|
Certificates & Approvals
| Quality certificate, Material certificate 3.1, Surface finish certificate, ATEX/IECEx/FM certificate, FDA/USP Class IV |
Sensor signal in air at 25 °C (77 °F)
| 50-110 nA |
Detection limit
| 6 ppb |
Measurement Range
| 0-100% O2 saturation |
Sterilizable/Autoclavable
| Yes |
Residual signal in oxygen free media
| <0.1% of signal in ambient air |
Measured Pressure Resistance
| 2.9 psi – 87 psi (0.2 bar – 6 bar) |
Measuring Temp. Range
| 32 °F – 176 °F (0 °C – 80 °C) |