The most common sources of interference are air bubbles at the membrane surface, depleted electrolyte, and an expired membrane, all of which cause unstable readings or drift in the measurement. When your readings become unreliable, these are the first things to check.
Flow rate also affects measurement. If flow past your sensor is too low, a depletion zone forms at the membrane surface where ozone is consumed faster than it is replenished, causing readings to appear lower than the actual process concentration. The recommended flow rate for the pureO3 flow housing is 200 to 500 mL/min.
Keep your sample line as short as possible and use stainless steel or an ozone-compatible polymer. Ozone decays rapidly, and a long sample line introduces measurement delay and concentration loss before the sample even reaches your sensor.