Need assistance?
Our team is here to achieve your goals. Speak with our experts.

6000TOCi 传感器, 100 - 240 VAC 50 - 60 Hz

实时总有机碳控制.6000TOCi在线传感器为有机污染物检测提供了真正的持续测量。6000TOCi可最快速地对TOC变化作出响应,是需要快速检测TOC变化的所有纯水应用中的理想之选。

实时TOC浓度监测

拥有最快的响应时间,并且每秒更新测量数值连续监测TOC分析浓度,能够准确无误地控制您的水系统。

决定性的数据满足合规性

6000TOCi采用成熟的紫外线氧化技术和高精度电导率传感器,提供一致、准确的有机物检测。

经过验证的系统性能

6000TOCi 具有高级实时诊断功能,让您洞悉所需一切信息,确保您的TOC分析仪始终保持更佳性能。

物料号: 30472150

6000TOCi在线TOC分析仪
拨打电话询价
规格 - 6000TOCi 传感器, 100 - 240 VAC 50 - 60 Hz
测量范围
0.05 - 2000 ppbC (µgC/L)
测量范围
0.05 ppb – 2,000 ppb
响应时间
60 s
检测极限
0.025 ppb
压力范围
0.3 bar – 13.6 bar
温度准确度(±)
± 0.25°C
分析时间
连续运行
检测限
0.025 ppbC
流速
> 8.5 mL/min
维护
每6个月更换紫外灯管
重量
5 kg (11.0 lb)
重现性
± 0.05 ppbC < 5 ppbC, ±1.0% > 5 ppbC
温度传感器
Pt1000 RTD, Class A
外形尺寸HWD
11.9” [302.75 mm] W x 9” [229.8 mm] H x 5.7” [144.7 mm] D
在线或便携
在线
外壳材质
粉末涂层铝制后盖和聚苯乙烯树脂前盖。
初始响应时间
< 60 seconds
分辨率
0.001 ppbC (µgC/L)
电导率精度
± 2%, 0.02 – 20 μS/cm ± 3%, 20 –100 μS/cm
电极常数(准确度)
± 2%
智能传感器管理(ISM®)
适用

实时合规性

6000TOCi分析仪能够快速和持续地测量水系统中的TOC水平。与批次测量系统相比,6000TOCi向用户提供不会错过哪怕是一分钟TOC超限的实时数据。

稳定可靠的分析

6000TOCi采用成熟的紫外线氧化技术和更高精度的电导率传感器,测量精准且具有可重复性。请放心,您将拥有满足监管要求的决定性的数据。

清晰的TOC历史记录支持审计检查

具有实时测量功能的6000TOCi可详细记录变化趋势、历史与关键统计信息,以支持您的合规文件记录。您可以通过用于打印和数据记录的USB端口获取完整的传感器历史记录。

符合法规要求

6000TOCi符合全球所有主要药典(包括USP、EP、JP、ChP和IP)对TOC仪器的要求。通过实时TOC分析,控制情况一目了然,全球和内部水质要求都能轻松满足。

文件记录

样本

Product Brochure: 6000TOCi with ISM
Continuous monitoring provides a real-time data advantage over batch or non-continuous methods of total organic carbon (TOC) measurement. Unlike at-li...

技术规格

Technical Data: 6000TOCi with ISM
The 6000TOCi is the latest edition to the METTLER TOLEDO Thornton On-line TOC Sensors. It’s dependable and reliable design uses proven UV oxidation te...

单页样本

Data Sheet: 6000TOCi with ISM
The 6000TOCi analyzer provides rapid, continuous measurement of TOC levels in your water system. Versus batch systems that measure at intervals, 6000T...
Data Sheet: 6000TOCi Services
Expert system services and periodic replacement of critical system components, such as the UV lamp, ensures your TOC system is always operating effici...

Manuals

Operation Manual: 6000TOCi Sensor
The 6000TOCi sensor is a Total Organic Carbon sensor with ISM capabilities, designed to measure the concentration of organic substances in pure and ul...
Quick Setup Guide: 6000TOCi & M800 Transmitter
Mount the 6000TOCi in an appropriate location. A minimum clearance dimension of 18"(45.7 cm) is required on the left side of the sensor for UV lamp ch...

白皮书

关于简化法规的白皮书
关于制药用水中总有机碳(TOC)法规的白皮书。

附件

TOC泵模块

该8.5 ml/min泵模块采用准确的容积式柱塞泵为TOC传感器提供高度稳定的工艺水样流量计量,以确保可靠、一致的TOC测量性能。该泵可与6000TOCi配合使用。

物料号: 30472152

耗材

软件

索取您的报价
6000TOCi 传感器, 100 - 240 VAC 50 - 60 Hz
FDA