No Watering Down of Sensor Data
All pH sensor calibration data is stored on a sensor chip and automatically and instantly transferred to the titrator as soon as it is connected.
Maintenance Free Reference
Double-junction design with maintenance-free internal reference system enables reference electrolyte replacement and adaptation to test conditions.
No Interference
A gold-coated, electrically shielded inner tube prevents interference from electrostatic charges, reducing signal noise and pH reading drift.
Material No.: 51109506
Electrolyte flow through the fixed glass junction remains stable between 5 and 20 μL/h, guaranteeing accurate pH readings and minimal sample contamination by electrolytes.
Temperature-adjusted pH readings and end point titration can be performed thanks to the integrated Pt1000 temperature sensor.
LoT glass brings low membrane resistance and ensures extremely fast response time. Thin-walled spherical design reduces electrical resistance even further.
Measurement Type | Acid/Base Potentiometric |
pH measuring range | 0 – 11 |
Int. reference electrolyte | KCl gel |
Membrane glass | LoT spherical |
Connection | S12 |
Ext. diaphragm | Ground glass |
Reference System | ARGENTHAL™ |
Bridge electrolyte | 1 mol/L KCl |
Type of Sensor | pH (Glass Membrane) Sensors |
Int. diaphragm | Ceramic pin |
Shaft Material | Glass |
Temperature Range | 32 °F – 212 °F (0 °C – 100 °C) |
Diaphragm | Ground glass (fixed) |
Detection | Plug & Play Sensor Chip |
Product Line | Excellence Titrators Potentiometric Compact Titrators |