The InPro 6950iG/12/120 is an amperometric gas-phase oxygen sensor designed for trace oxygen measurement in CO2 recovery in breweries. This sensor uses Intelligent Sensor Management (ISM) diagnostics to reduce maintenance and downtime.
Accurate Trace Oxygen Measurement
The InPro 6950iG/12/120 provides an accuracy of ±5 ppm in CO2 recovery applications in breweries, where trace level accuracy is critical.
Save Money with CO2 Recovery
Using the InPro 6950iG/12/120 helps measure and capture CO2 from fermentation so you can reduce CO2 purchases for carbonation.
Real-Time Sensor Performance Monitoring
The InPro 6950iG/12/120 uses ISM technology to provide advanced analytics of sensor health, calibration needs, and remaining membrane lifetime.
Measurement Parameter | Gas-Phase Oxygen |
Measurement Technology | Amperometric Clark Electrode |
Measurement Range | 6 ppm - 50000ppm |
Accuracy | ≤ ± [1% +5 ppm] |
Response Time | T90 ≤ 30s (at 25 °C / 77 °F) |
Primary Media/Application | Chemical Processes Food/Beverage Production |
ISM (Digital) or Analog | ISM (Digital) |
Cable Connector | K8S |
Sensor Length | 120 mm |
Process Connection | Pg 13.5 |
Connector Design | Pg 13.9 |
O-Ring Material | Silicone |
Sensor Diameter | 12 mm |
Surface Roughness of Wetted Parts | N5/Ra 16 (Ra=0.4mm/16min) |
Wetted Materials | 316L Stainless Steel |
Wetted membrane material | PTFE |
Model Family | InPro6950iG |
Detection limit | 5 ppm |
Detection limit | 5 ppm |
Sensor signal in air at 25 °C (77 °F) | 2500...6000 nA |
Residual signal in oxygen free media | <0.025 % of signal in ambient air |
Temperature compensation | Automatic |
Measuring Temp. Range | 32 °F – 104 °F (0 °C – 40 °C) |
Operating Pressure (Bar) | 0.200 bar–9.000 bar |
Operating Pressure (PSI) | 2.9 to 130 psi absolute |
Mechanical Temp. Range | 23 °F – 249.8 °F (-5 °C – 121 °C) |
Design pressure (bar) | maximum 12 bar |
Design Pressure (PSI) | maximum 174 psi absolute |
Hazardous Area Approvals | ATEX: Ex ia IIC T6 / T5 / T4 / T3 Ga / Gb, Ex ia IIIC T69 °C / T81 °C / T109 °C / T161 °C Da / Db. FM: IS Cl. I, II, III, Div 1, GR ABCDEFG / T6 * |
Hygienic/Biocompatibility Approvals | EHEDG, FDA/USP Class VI |
Other certificates | METTLER TOLEDO Quality Certificate, Material certificate 3.1, N5/Ra 17 |
Operating Range | 5 ppm – 50,000 ppm |
Mechanical Temperature Resistance (Celsius) | -5 °C – 121 °C |
Connection | K8S ISM |
Measured Pressure Resistance | 0.2 bar – 9 bar |
Mechanical Temperature Resistance (Fahrenheit) | 23 °F – 249.8 °F |
Accurate at Low Oxygen Concentrations
The InPro 6950iG/12/120 offers reliable measurement between 5 and 50,000 ppm oxygen at an accuracy of ±5 ppm. It gives you the low concentration measurements you need during CO2 recovery and separation to save you costs on CO2 purchases.
Fast Sensor Replacement for More Uptime
The InPro 6950iG/12/120 offers an easy-to-change membrane body and ISM Plug and Measure technology. Plug and Measure allows a sensor to be polarized and calibrated away from the process under lab conditions, and then inserted directly into the process.
Meets Your Installation Needs
The InPro 6950iG/12/120 offers a 120 mm length suitable for short to medium length installations. The membrane body includes a silicone O-ring and 316L stainless steel wetted parts, making it appropriate for CO2 recovery in the brewery industry.
Plan Maintenance with ISM Diagnostics
ISM technology in the InPro 6950iG/12/120 monitors sensor stress to provide a prediction for upcoming calibration needs. It also tracks the electrolyte level to identify evaporation and leakage, and informs you when refilling is required.