InPro 2000 120/PT100/9848 Refillable Analog pH Sensor

pH Sensor InPro 2000 120/PT100/9848

Refillable pH Probe with Friscolyt

The InPro 2000 120/PT100/9848 is a refillable analog pH probe with an integrated PT100 temperature sensor for demanding applications. This 120 mm refillable pH sensor with Friscolyt electrolyte is a reliable pH measurement solution.

For Use in Extreme Conditions
The InPro 2000 120/PT100/9848 uses a pressurizable reference electrolyte to ensure measurement stability and low maintenance in harsh applications.

Refillable Sensor with Long Lifetime
This pH sensor uses a refillable Friscolyt electrolyte to ensure reliable results over a long lifetime in media with proteins/organic solvents.

Watertight connection
The IP 68 watertight cable connection provides excellent signal transmission in all environments.

pH Sensor InPro 2000 120/PT100/9848
Specifikationer - pH Sensor InPro 2000 120/PT100/9848
Analog eller ISM Analog
Mätningsparametrar pH
Typ av pH-membranglas High Alkali glass (HA)
Enkel eller dubbel förgrening Single Junction Refillable
Sensorlängd 120 mm
Temperatursignal PT100
Anslutningstyp VP
Lösningsjord/redoxmetall None
Elektrolyt Friscolyt
pH range 0-14 pH
Tryckresistens (bar) 0 to 6 barg (in presurized housing)
Tryckresistens (psi) 0 to 87 psig (in presurized housing)
Reference system Argenthal with silver-ion trap
Steriliserbar Yes
Autoklaverbar Yes
Membran Ceramic
ATEX Certification SEV 14 ATEX 0168 X
Ex ia IIC T6/T5/T4/T3 Ga/Gb
FM approval IS/ I, II, III /1/ABCDEFG /T6
IECEx Certificate of Conformity IECEx SEV 14.0025X
Ex ia IIC T6/T5/T4/T3 Ga/Gb
Mättemperaturområde (imp) 32 to 284° F
Mättemperaturområde (metriskt) 0 to 140° C
Materialnummer 52001435

Features and Benefits


Reliable in Demanding Applications
The InPro 2000 120/PT100/9848 refillable pH sensor uses a pressurizable reference electrolyte that ensures permanent self-cleaning of the diaphragm. This leads to long-term pH measurement stability and low maintenance in harsh applications.
 
Reliability in Challenging Media
This refillable pH sensor uses Friscolyt electrolyte. This electrolyte is designed to provide optimal results and an extended lifetime in media with high protein or organic solvent concentrations. It's also well suited for low temperature media.
 
Meets Your Installation Requirements
The InPro 2000 120/PT100/9848 has an integrated PT100 temperature signal for reliable temperature compensation. With a 120 mm length, it is optimal when short installation lengths are needed.
 
Accurate, Reliable Measurement System 
The InPro 2000 is designed with long-term accuracy and reliability in mind. It is sterilizable and autoclavable, has a silver-ion trap to prevent sulfide poisoning of the junction, and is ATEX and FM certified for installation in hazardous areas.

Dokumentation

Program

Application Note: Improve Your Pulp & Paper Process Further with ISM Technology
Application Note - Analítica de procesos: “Nota de aplicación”
Easy ISM Sensor Management in Breweries
Application Note - iSense ISM Asset Suite the professional software tool to implement innovative and efficient maintenance concepts with pre-calibrate...
pH Control in Sugar Refineries
The process of refining sugar from cane or beets involves the following steps: washing, grinding, juicing, liming, carbonation, filtration, sulfitatio...
Få din offert
pH Sensor InPro 2000 120/PT100/9848

Tillbehör

Få din offert
pH Sensor InPro 2000 120/PT100/9848
Thank you for visiting www.mt.com. We have tried to optimize your experience while on the site, but we noticed that you are using an older version of a web browser. We would like to let you know that some features on the site may not be available or may not work as nicely as they would on a newer browser version. If you would like to take full advantage of the site, please update your web browser to help improve your experience while browsing www.mt.com.