Measurement Parameter
| Oxygen |
Measurement Technology
| Optical fluorescence quenching |
Accuracy
| ± (1 % of the reading + 8 ppb) |
Response Time
| T98 = < 20 s |
Primary Media/Application
| Food and Beverage |
ISM (Digital) or Analog
| ISM (Digital) |
Wetted Metallic Material
| 316L stainless steel |
Surface Roughnes of Wetted Material
| N5/Ra16 (Ra=0.4μm/16μin) |
Sensor Diameter
| 12mm |
O-Ring Material
| EPDM |
Process Connection
| Pg 13.5 |
Sensor Length
| 120 mm |
Wetted membrane material
| Silicone |
Connector Design
| Straight |
Model Family
| InPro 6900i optical series |
Autoclavable/Sterilizable
| Yes |
CIP resistant
| Yes |
Mechanical Temp. Range
| 23 °F – 250 °F (-5 °C – 121 °C) |
Operating Pressure (Bar)
| 0.2–12 Bar |
Operating Pressure (PSI)
| 02.9–174 PSI |
Design Pressure
| 174 psi (12 bar) |
Digital Integration
| RS485 - ISM® |
Cable Connector
| M12 5-Pin |
Hygienic/Biocompatibility Approvals
| FDA/USP Class 7 |
Certificates & Approvals
| Quality certificate, Material certificate 3.1, Surface finish certificate, FDA/USP Class IV |
Measurement Range
| 0- 25 ppm saturation |
Residual signal in O2 free media
| <0.025 of signal in ambient air |
Operating Range
| 8 ppb - 60% O2 sat. |
Sterilizable/Autoclavable
| Sterilizable |
Measured Pressure Resistance
| 3 bar – 12 bar |
Measuring Temp. Range
| 23 °F – 104 °F (-5 °C – 40 °C) |
Detection limit
| 8 ppb |
Connection
| 5-pin |