Effects of poor maintenance and extended use of titration sensors.

 

What are the reasons that

a. A titration curve becomes noisy
b. the inflection point in an equivalence point titration becomes flat
c. the results are too low

Possible reasons:

Poor maintenance and extended use of titration sensors, insufficient sensor response.

Recommendations:

Poor maintenance and use of sensors that can no longer meet calibration or kinetic sensor test requirements can result in multiple errors when performing the titration. The degradation of a sensor can result in a slow response time which can be recognized as a high drift value during the kinetic sensor test that is available as ‘Sensor Check’ routine in the titrator for pH sensors.

Another result of a poorly maintained or old sensor is that it may no longer properly associate millivolt readings to the proper concentration. With calibrated sensors, this is observed as a decrease in the slope outside of accepted ranges. With redox and argentometric sensors, it is recommended to assess the potential (mV) change at the equivalence point. The potential change should show at least 100 mV at the inflection.