Single-use Sensors
Single-use Sensors

Single-use Sensors for Bag Manufacturers

pH and Dissolved Oxygen for stirred Single-use Reactors (> 20 L)

Designed for Single-Use Bag Manufacturers
Easy and Safe Installation and Integration
Reliable, Straightforward Use in Biocontrollers

Produkter og detaljer

 
Produkter og detaljer
Filter:
Sett filter
Clear All
 
Kort beskrivelse
Måleområde
Se detaljer
Kort beskrivelseConventional pH sensor with pH sensitive membrane glass
Måleområde pH 3-10
FøleelementerGlass pH and Reference Electrodes, Lead-Free
Integrert temperatursondePt1000
Nullpunkt7.20 ±0.25 pH
Kalibrering1-point process
SkaftmaterialerHigh density polyethylene, USP 88 class VI
O-ring materialEPDM, USP 88 class VI
ProsessforholdEldon James 1' port disc, with retainer
Kan steriliseres med gammastråler25…45 kGy
Operating Temperature5…60°C (41…140°F)
Storage temperature5…25°C (41…77°F)
Mekanisk trykkmotstand"2 barg / 40 °C (29 psig / 104 °F), 1 barg / 60 °C (15 psig / 140 °F)"
GodkjenningerUSP 88 class VI for all wetted polymer parts, factory calibration
Se detaljer
Kort beskrivelseOptical dissolved oxygen sensor
Måleområde 0…250% air
FøleelementerOptical spot (Fluorescensce quenching)
Integrert temperatursondePt1000
Kalibrering1-point air
SkaftmaterialerHigh density polyethylene, USP 88 class VI
O-ring materialEPDM, USP 88 class VI
ProsessforholdEldon James 1' port disc
Kan steriliseres med gammastråler25…45 kGy
Operating Temperature5…60°C (41…140°F)
Storage temperature5…25°C (41…77°F)
Mekanisk trykkmotstand"2 barg / 40 °C (29 psig / 104 °F), 1 barg / 60 °C (15 psig / 140 °F)"
GodkjenningerUSP 88 class VI for all wetted polymer parts
Comparison

Dokumentasjon

Application Note

Application Note: Differences in Operation Between Analytical Multi-Use and Single-Use Sensors in Bioreactors
This application note provides information on the important trend towards the deployment of single-use bioreactors (SUBs) in the bioprocess industry....

Teknisk spesifikasjon

Technical Data: InSUS x07 Sensor Series
The InSUS™ x07 sensor series for single-use technology bioprocesses cover the measurement parameters pH and dissolved oxygen (DO).

Instruction Manuals

Instruction Manual: InSUS Head for Optical Single-Use O₂ Sensors
These operating instructions contain all the information needed for safe and proper use of the optical sensor head.
Instruction Manual: Information for InSUS fiber optic cable
The enclosed adapter must be used, if the fiber optic cable is connected to sensors of the type InSUS 607. Mount the adapter – as shown in the photos...
M300 / M400 2G Quick Calibration Guide For pH Single-Use Sensors Type InSUS 307
METTLER TOLEDO InSUS™ pH sensors are analog, gamma irradiation-sterilizable single-use pH sensors with an integrated Pt 1000 temperature probe.

Service

+47 22 30 44 90
Kontakt service
Thank you for visiting www.mt.com. We have tried to optimize your experience while on the site, but we noticed that you are using an older version of a web browser. We would like to let you know that some features on the site may not be available or may not work as nicely as they would on a newer browser version. If you would like to take full advantage of the site, please update your web browser to help improve your experience while browsing www.mt.com.