Operating Instructions: Optical O₂ Simulator

With the Optical O₂ Simulator it is possible to simulate the output of different digital optical oxygen sensors. The simulator is connected to the transmitter instead of the real sensor to verify connections are correct, and to simulate the behavior of the transmitter in different measurement conditions and in case of a suspected sensors error.

Depending on the chosen condition mode, the simulator works as if it is measuring in air or in an oxygen-free medium. 

As soon as the simulator is connected to the transmitter, the transmitter will show pre-defined data such as DLI, ACT, operating hours, autoclaving cycles, SIP and CIP of the selected sensor.

Operating Instructions: Optical O₂ Simulator

With the Optical O₂ Simulator it is possible to simulate the output of different digital optical oxygen sensors. The simulator is connected to the transmitter instead of the real sensor to verify connections are correct, and to simulate the behavior of the transmitter in different measurement conditions and in case of a suspected sensors error.