Titration calibration is basically the verification of the technical specification of a titrator. For example, the burette drive stroke length significantly impacts the dosing accuracy, which is crucial to achieving accurate and reliable titration results. We use a certified micrometer and measure the stroke length at different levels (in accordance with ISO 8655) to ensure dosing accuracy over the entire dispensing range/burette volume. The burette volume is calibrated separately, according to ISO 8655 part 3, to ensure truly interchangeable burettes.
The accuracy of the sensor input is indispensable for proper potential measurements (e.g., mV/pH) to monitor the titration reaction and the correct EP/EQP evaluation. For calibration, we are using certified resistors and digital multimeter to ensure measurements are within specifications.
For the InMotion™ KF Oven Autosampler, the oven temperature accuracy and stability and gas flow rate are calibrated using a certified temperature sensor and gas flow meter.
The certificate includes a clear pass/fail decision based on maximum permissible errors from technical specifications and provides full traceability of all measurement results and tools used for calibration.