Amperometric Sensor for Gas-Phase Oxygen
Amperometric sensor

Amperometric Sensor for Gas-Phase Oxygen

Amperometric sensors & probes for oxygen gas measurement without the need for conditioning systems

Amperometric Sensor for Gas-Phase Oxygen

In-line amperometric sensors (also called polarographic sensors) for gas-phase oxygen measurement in chemical processes combine high operational up...

In-line amperometric sensors (also called polarographic sensors) for gas-phase oxygen measurement in chemical processes combine high operational uptime with excellent performance. Without the need for an expensive gas sample conditioning system, METTLER TOLEDO's gas-phase oxygen sensors can be installed directly in the process and sensor maintenance can be performed without process interruption.

METTLER TOLEDO offers a unique, easy-to-use and reliable solution for challenging applications such as N2 blanketing, inertization and off-gas monitoring in hazardous or non-hazardous applications.

Oxygen sensors with integrated Intelligent Sensor Management (ISM) functionality feature advanced diagnostics, and Plug and Measure for fast sensor start-up.


Amperometric Oxygen Gas Sensor
Unaffected by Environmental Challenges

Better Process Control With Trace Measurement

Products and Specs

 
Products and Specs
Filter:
Set filter
Clear All
 
Analog or ISM
Operating range
Length
O-ring material
Membrane Body Wetted Parts Material
See details
Analog or ISMAnalog
Operating range0.1 Vol-% O₂ to 100 Vol-% O₂
Length120mm
O-ring materialSilicone
Membrane Body Wetted Parts Material316L Stainless Steel
Accuracy≤ ±[1% +0.1 Vol-%]
90% response time at 25°C (77°F)≤ 20 s (N2@15 Vol-% O₂)
Sensor Signal in air at 25°C (77°F)50 to 110 nA
Cable ConnectionStraight VP
Measuring pressure resistance0.2 to 9 bar (2.9 to 130 psi absolute)
CertificatesSEV 14 ATEX 0169 X
Ex ia IIC T6/T5/T4/T3 Ga/Gb
Ex ia IIIC T69°C/T81°C/T109°C/ T161°C Da/Db
IECEx Certificate of ConformityIECEx SEV 14.0026X
Ex ia IIC T6/T5/T4/T3 Ga/Gb
Ex ia IIIC T69°C/T81°C/T109°C/ T161°C Da/Db
See details
Analog or ISMAnalog
Operating range0.1 Vol-% O₂ to 100 Vol-% O₂
Length220mm
O-ring materialSilicone
Membrane Body Wetted Parts Material316L Stainless Steel
Accuracy≤ ±[1% +0.1 Vol-%]
90% response time at 25°C (77°F)≤ 20 s (N₂@15 Vol-% O₂)
Sensor Signal in air at 25°C (77°F)50 to 110 nA
Cable ConnectionStraight VP
Measuring pressure resistance0.2 to 9 bar (2.9 to 130 psi absolute)
CertificatesSEV 14 ATEX 0169 X
Ex ia IIC T6/T5/T4/T3 Ga/Gb
Ex ia IIIC T69°C/T81°C/T109°C/ T161°C Da/Db
IECEx Certificate of ConformityIECEx SEV 14.0026X
Ex ia IIC T6/T5/T4/T3 Ga/Gb
Ex ia IIIC T69°C/T81°C/T109°C/ T161°C Da/Db
See details
Analog or ISMAnalog
Operating range0.1 Vol-% O₂ to 100 Vol-% O₂
Length120mm
O-ring materialKalrez®
Membrane Body Wetted Parts Material316L Stainless Steel
Accuracy≤ ±[1% +0.1 Vol-%]
90% response time at 25°C (77°F)≤ 20 s (N₂@15 Vol-% O₂)
Sensor Signal in air at 25°C (77°F)50 to 110 nA
Cable ConnectionStraight VP
Measuring pressure resistance0.2 to 9 bar (2.9 to 130 psi absolute)
CertificatesSEV 14 ATEX 0169 X
Ex ia IIC T6/T5/T4/T3 Ga/Gb
Ex ia IIIC T69°C/T81°C/T109°C/ T161°C Da/Db
IECEx Certificate of ConformityIECEx SEV 14.0026X
Ex ia IIC T6/T5/T4/T3 Ga/Gb
Ex ia IIIC T69°C/T81°C/T109°C/ T161°C Da/Db
See details
Analog or ISMAnalog
Operating range0.1 Vol-% O₂ to 100 Vol-% O₂
Length220mm
O-ring materialKalrez®
Membrane Body Wetted Parts Material316L Stainless Steel
Accuracy≤ ±[1% +0.1 Vol-%]
90% response time at 25°C (77°F)≤ 20 s (N₂@15 Vol-% O₂)
Sensor Signal in air at 25°C (77°F)50 to 110 nA
Cable ConnectionStraight VP
Measuring pressure resistance0.2 to 9 bar (2.9 to 130 psi absolute)
CertificatesSEV 14 ATEX 0169 X
Ex ia IIC T6/T5/T4/T3 Ga/Gb
Ex ia IIIC T69°C/T81°C/T109°C/ T161°C Da/Db
IECEx Certificate of ConformityIECEx SEV 14.0026X
Ex ia IIC T6/T5/T4/T3 Ga/Gb
Ex ia IIIC T69°C/T81°C/T109°C/ T161°C Da/Db
See details
Analog or ISMAnalog
Operating range0.1 Vol-% O₂ to 100 Vol-% O₂
Length120mm
O-ring materialKalrez®
Membrane Body Wetted Parts MaterialC22 Hastelloy
Accuracy≤ ±[1% +0.1 Vol-%]
90% response time at 25°C (77°F)≤ 20 s (N₂@15 Vol-% O₂)
Sensor Signal in air at 25°C (77°F)50 to 110 nA
Cable ConnectionStraight VP
Measuring pressure resistance0.2 to 9 bar (2.9 to 130 psi absolute)
CertificatesSEV 14 ATEX 0169 X
Ex ia IIC T6/T5/T4/T3 Ga/Gb
Ex ia IIIC T69°C/T81°C/T109°C/ T161°C Da/Db
IECEx Certificate of ConformityIECEx SEV 14.0026X
Ex ia IIC T6/T5/T4/T3 Ga/Gb
Ex ia IIIC T69°C/T81°C/T109°C/ T161°C Da/Db
See details
Analog or ISMAnalog
Operating range0.1 Vol-% O₂ to 100 Vol-% O₂
Length220mm
O-ring materialKalrez®
Membrane Body Wetted Parts MaterialC22 Hastelloy
Accuracy≤ ±[1% +0.1 Vol-%]
90% response time at 25°C (77°F)≤ 20 s (N₂@15 Vol-% O₂)
Sensor Signal in air at 25°C (77°F)50 to 110 nA
Cable ConnectionStraight VP
Measuring pressure resistance0.2 to 9 bar (2.9 to 130 psi absolute)
CertificatesSEV 14 ATEX 0169 X
Ex ia IIC T6/T5/T4/T3 Ga/Gb
Ex ia IIIC T69°C/T81°C/T109°C/ T161°C Da/Db
IECEx Certificate of ConformityIECEx SEV 14.0026X
Ex ia IIC T6/T5/T4/T3 Ga/Gb
Ex ia IIIC T69°C/T81°C/T109°C/ T161°C Da/Db
See details
Analog or ISMISM
Operating range0.1 Vol-% O₂ to 100 Vol-% O₂
Length120mm
O-ring materialSilicone
Membrane Body Wetted Parts Material316L Stainless Steel
Accuracy≤ ±[1% +0.1 Vol-%]
90% response time at 25°C (77°F)≤ 20 s (N₂@15 Vol-% O₂)
Sensor Signal in air at 25°C (77°F)50 to 110 nA
Cable ConnectionStraight K8S
Measuring pressure resistance0.2 to 9 bar (2.9 to 130 psi absolute)
CertificatesSEV 14 ATEX 0169 X
Ex ia IIC T6/T5/T4/T3 Ga/Gb
Ex ia IIIC T69°C/T81°C/T109°C/ T161°C Da/Db
IECEx Certificate of ConformityIECEx SEV 14.0026X
Ex ia IIC T6/T5/T4/T3 Ga/Gb
Ex ia IIIC T69°C/T81°C/T109°C/ T161°C Da/Db
See details
Analog or ISMISM
Operating range0.1 Vol-% O₂ to 100 Vol-% O₂
Length220mm
O-ring materialSilicone
Membrane Body Wetted Parts Material316L Stainless Steel
Accuracy≤ ±[1% +0.1 Vol-%]
90% response time at 25°C (77°F)≤ 20 s (N₂@15 Vol-% O₂)
Sensor Signal in air at 25°C (77°F)50 to 110 nA
Cable ConnectionStraight K8S
Measuring pressure resistance0.2 to 9 bar (2.9 to 130 psi absolute)
CertificatesSEV 14 ATEX 0169 X
Ex ia IIC T6/T5/T4/T3 Ga/Gb
Ex ia IIIC T69°C/T81°C/T109°C/ T161°C Da/Db
IECEx Certificate of ConformityIECEx SEV 14.0026X
Ex ia IIC T6/T5/T4/T3 Ga/Gb
Ex ia IIIC T69°C/T81°C/T109°C/ T161°C Da/Db
See details
Analog or ISMISM
Operating range0.1 Vol-% O₂ to 100 Vol-% O₂
Length120mm
O-ring materialKalrez®
Membrane Body Wetted Parts Material316L Stainless Steel
Accuracy≤ ±[1% +0.1 Vol-%]
90% response time at 25°C (77°F)≤ 20 s (N₂@15 Vol-% O₂)
Sensor Signal in air at 25°C (77°F)50 to 110 nA
Cable ConnectionStraight K8S
Measuring pressure resistance0.2 to 9 bar (2.9 to 130 psi absolute)
CertificatesSEV 14 ATEX 0169 X
Ex ia IIC T6/T5/T4/T3 Ga/Gb
Ex ia IIIC T69°C/T81°C/T109°C/ T161°C Da/Db
IECEx Certificate of ConformityIECEx SEV 14.0026X
Ex ia IIC T6/T5/T4/T3 Ga/Gb
Ex ia IIIC T69°C/T81°C/T109°C/ T161°C Da/Db
See details
Analog or ISMISM
Operating range0.1 Vol-% O₂ to 100 Vol-% O₂
Length220mm
O-ring materialKalrez®
Membrane Body Wetted Parts Material316L Stainless Steel
Accuracy≤ ±[1% +0.1 Vol-%]
90% response time at 25°C (77°F)≤ 20 s (N₂@15 Vol-% O₂)
Sensor Signal in air at 25°C (77°F)50 to 110 nA
Cable ConnectionStraight K8S
Measuring pressure resistance0.2 to 9 bar (2.9 to 130 psi absolute)
CertificatesSEV 14 ATEX 0169 X
Ex ia IIC T6/T5/T4/T3 Ga/Gb
Ex ia IIIC T69°C/T81°C/T109°C/ T161°C Da/Db
IECEx Certificate of ConformityIECEx SEV 14.0026X
Ex ia IIC T6/T5/T4/T3 Ga/Gb
Ex ia IIIC T69°C/T81°C/T109°C/ T161°C Da/Db
See details
Analog or ISMISM
Operating range0.1 Vol-% O₂ to 100 Vol-% O₂
Length120mm
O-ring materialKalrez®
Membrane Body Wetted Parts Material316L Stainless Steel
Accuracy≤ ±[1% +0.1 Vol-%]
90% response time at 25°C (77°F)≤ 20 s (N₂@15 Vol-% O₂)
Sensor Signal in air at 25°C (77°F)50 to 110 nA
Cable ConnectionStraight K8S
Measuring pressure resistance0.2 to 9 bar (2.9 to 130 psi absolute)
CertificatesSEV 14 ATEX 0169 X
Ex ia IIC T6/T5/T4/T3 Ga/Gb
Ex ia IIIC T69°C/T81°C/T109°C/ T161°C Da/Db
IECEx Certificate of ConformityIECEx SEV 14.0026X
Ex ia IIC T6/T5/T4/T3 Ga/Gb
Ex ia IIIC T69°C/T81°C/T109°C/ T161°C Da/Db
See details
Analog or ISMISM
Operating range0.1 Vol-% O₂ to 100 Vol-% O₂
Length220mm
O-ring materialKalrez®
Membrane Body Wetted Parts MaterialC22 Hastelloy
Accuracy≤ ±[1% +0.1 Vol-%]
90% response time at 25°C (77°F)≤ 20 s (N₂@15 Vol-% O₂)
Sensor Signal in air at 25°C (77°F)50 to 110 nA
Cable ConnectionStraight K8S
Measuring pressure resistance0.2 to 9 bar (2.9 to 130 psi absolute)
CertificatesSEV 14 ATEX 0169 X
Ex ia IIC T6/T5/T4/T3 Ga/Gb
Ex ia IIIC T69°C/T81°C/T109°C/ T161°C Da/Db
IECEx Certificate of ConformityIECEx SEV 14.0026X
Ex ia IIC T6/T5/T4/T3 Ga/Gb
Ex ia IIIC T69°C/T81°C/T109°C/ T161°C Da/Db
See details
Analog or ISMISM
Operating range50 ppm to 60 Vol-% O₂
Length120mm
O-ring materialSilicone
Membrane Body Wetted Parts Material316L Stainless Steel
Accuracy≤ ± [1% +50 ppm]
90% response time at 25°C (77°F)≤ 20 s (N₂@15 Vol-% O₂)
Sensor Signal in air at 25°C (77°F)250 to 500 nA
Cable ConnectionStraight K8S
Measuring pressure resistance0.2 to 9 bar (2.9 to 130 psi absolute)
CertificatesSEV 14 ATEX 0169 X
Ex ia IIC T6/T5/T4/T3 Ga/Gb
Ex ia IIIC T69°C/T81°C/T109°C/ T161°C Da/Db
IECEx Certificate of ConformityIECEx SEV 14.0026X
Ex ia IIC T6/T5/T4/T3 Ga/Gb
Ex ia IIIC T69°C/T81°C/T109°C/ T161°C Da/Db
See details
Analog or ISMISM
Operating range50 ppm to 60 Vol-% O₂
Length220mm
O-ring materialSilicone
Membrane Body Wetted Parts Material316L Stainless Steel
Accuracy≤ ± [1% +50 ppm]
90% response time at 25°C (77°F)≤ 20 s (N₂@15 Vol-% O₂)
Sensor Signal in air at 25°C (77°F)250 to 500 nA
Cable ConnectionStraight K8S
Measuring pressure resistance0.2 to 9 bar (2.9 to 130 psi absolute)
CertificatesSEV 14 ATEX 0169 X
Ex ia IIC T6/T5/T4/T3 Ga/Gb
Ex ia IIIC T69°C/T81°C/T109°C/ T161°C Da/Db
IECEx Certificate of ConformityIECEx SEV 14.0026X
Ex ia IIC T6/T5/T4/T3 Ga/Gb
Ex ia IIIC T69°C/T81°C/T109°C/ T161°C Da/Db
See details
Analog or ISMISM
Operating range50 ppm to 60 Vol-% O₂
Length120mm
O-ring materialKalrez®
Membrane Body Wetted Parts Material316L Stainless Steel
Accuracy≤ ± [1% +50 ppm]
90% response time at 25°C (77°F)≤ 20 s (N₂@15 Vol-% O₂)
Sensor Signal in air at 25°C (77°F)250 to 500 nA
Cable ConnectionStraight K8S
Measuring pressure resistance0.2 to 9 bar (2.9 to 130 psi absolute)
CertificatesSEV 14 ATEX 0169 X
Ex ia IIC T6/T5/T4/T3 Ga/Gb
Ex ia IIIC T69°C/T81°C/T109°C/ T161°C Da/Db
IECEx Certificate of ConformityIECEx SEV 14.0026X
Ex ia IIC T6/T5/T4/T3 Ga/Gb
Ex ia IIIC T69°C/T81°C/T109°C/ T161°C Da/Db
See details
Analog or ISMISM
Operating range50 ppm to 60 Vol-% O₂
Length220mm
O-ring materialKalrez®
Membrane Body Wetted Parts Material316L Stainless Steel
Accuracy≤ ± [1% +50 ppm]
90% response time at 25°C (77°F)≤ 20 s (N₂@15 Vol-% O₂)
Sensor Signal in air at 25°C (77°F)250 to 500 nA
Cable ConnectionStraight K8S
Measuring pressure resistance0.2 to 9 bar (2.9 to 130 psi absolute)
CertificatesSEV 14 ATEX 0169 X
Ex ia IIC T6/T5/T4/T3 Ga/Gb
Ex ia IIIC T69°C/T81°C/T109°C/ T161°C Da/Db
IECEx Certificate of ConformityIECEx SEV 14.0026X
Ex ia IIC T6/T5/T4/T3 Ga/Gb
Ex ia IIIC T69°C/T81°C/T109°C/ T161°C Da/Db
See details
Analog or ISMISM
Operating range5 ppm to 50,000 ppm
Length120mm
O-ring materialSilicone
Membrane Body Wetted Parts Material316L Stainless Steel
Accuracy≤ ± [1% +5 ppm]
90% response time at 25°C (77°F)≤ 20 s (N₂@15 Vol-% O₂)
Sensor Signal in air at 25°C (77°F)2500 to 6500 nA
Cable ConnectionStraight K8S
Measuring pressure resistance0.2 to 9 bar (2.9 to 130 psi absolute)
CertificatesSEV 14 ATEX 0169 X
Ex ia IIC T6/T5/T4/T3 Ga/Gb
Ex ia IIIC T69°C/T81°C/T109°C/ T161°C Da/Db
IECEx Certificate of ConformityIECEx SEV 14.0026X
Ex ia IIC T6/T5/T4/T3 Ga/Gb
Ex ia IIIC T69°C/T81°C/T109°C/ T161°C Da/Db
See details
Analog or ISMISM
Operating range5 ppm to 50,000 ppm
Length220mm
O-ring materialSilicone
Membrane Body Wetted Parts Material316L Stainless Steel
Accuracy≤ ± [1% +5 ppm]
90% response time at 25°C (77°F)≤ 20 s (N₂@15 Vol-% O₂)
Sensor Signal in air at 25°C (77°F)2500 to 6500 nA
Cable ConnectionStraight K8S
Measuring pressure resistance0.2 to 9 bar (2.9 to 130 psi absolute)
CertificatesSEV 14 ATEX 0169 X
Ex ia IIC T6/T5/T4/T3 Ga/Gb
Ex ia IIIC T69°C/T81°C/T109°C/ T161°C Da/Db
IECEx Certificate of ConformityIECEx SEV 14.0026X
Ex ia IIC T6/T5/T4/T3 Ga/Gb
Ex ia IIIC T69°C/T81°C/T109°C/ T161°C Da/Db
Comparison

Documentation

Case Studies

Protection from Gas Explosion inertization During Centrifugation
Comi Condor are one of the leading suppliers of vertical continuous and horizontal centrifuges for the fine chemicals industry. It has selected METTLE...
Flare Safety
This case study explains how Ningbo LG Yongxing Chemical Co., Ltd in China was using a paramagnetic analyzer for oxygen measurement in a flare system....
Fermentation processes
Sustainable biomass products are being increasingly adopted worldwide. To ensure reliability in fermentation processes in a renewable oil production p...
Centrifuge Safety and Productivity Case Study
Comi Condor are one of the leading suppliers of vertical continuous and horizontal centrifuges for the fine chemicals industry. It has selected METTLE...

Applications

Protection From Gas Explosion
One of the top-ten DMT producers was faced with periodic deficiencies of their oxygen measurement system for centrifuge inertization processes. The us...
In Situ Gas Analysis – Wherever It Is Required
The GPro 500 can also be used with an existing extraction system, which reduces the need for re-engineering and means expensive extraction systems are...
Ensuring Safety in Reactor and Centrifuge
Critical reaction and extraction stage in an API production process can be ensured safe with inline gas-phase oxygen sensing technology which eliminat...
Application Note: BioPharma - Optical Oxygen Systems
To enable optimal growth conditions for high yield and quality in biotech processes, control of the oxygen level is essential. Maintenance or replacem...
DO Measurement in Primary and Secondary Wastewater Treatment
Biological industrial wastewater treatment Biological processes are being more and more used to treat wastewater effluents, instead of chemical and m...

Services

Uptime
Support & Repair
Performance
Maintenance & Optimization
Expertise
Training & Consulting
Thank you for visiting www.mt.com. We have tried to optimize your experience while on the site, but we noticed that you are using an older version of a web browser. We would like to let you know that some features on the site may not be available or may not work as nicely as they would on a newer browser version. If you would like to take full advantage of the site, please update your web browser to help improve your experience while browsing www.mt.com.