PT4805 DPAS Redox Probe for Biotech & Biopharmaceuticals

Pt4805-DPAS-SC-K8S Redox/ORP Probe

Autoclavable gel-filled Redox-electrode for the biotech industry

Pt4805 is the Redox version of DPAS, i.e. a gel-filled, low-maintenance electrode, for in-line measuremernts of oxidation reduction potential (ORP) in biotech applications in small to medium-sized bioreactors at moderate pressure.
المواصفات - Pt4805-DPAS-SC-K8S Redox/ORP Probe
Short description Sterilizable gel-filled electrolyte Redox-electrode
pH-range mV
Temperature range 0...130 °C (32...266 °F)
Pressure resistance (bar) up to 2.5 bar overpressure
Pressure resistance (psi) up to 36 psi overpressure
Diaphragm ceramic
Number of diaphragms 1
Reference system Argenthal system
Silver-ion trap yes
Sterilizable yes
Autoclavable yes
ISM no
Model Family DPAS
Analog or ISM Analog
Measurement parameter ORP
Sensor Length 120mm
Electrolyte Gel
رقم/أرقام المواد 59904198
Features and Benefits
Low-maintenance
  • Pt4805 electrode is filled with gel-reference electrolyte which is pre-pressurized to 2.5 bar (36 psi). The internal over-pressure ensures a gradual electrolyte outflow efficiently cleaning the diaphragm and preventing influx of process media into the reference electrolyte.
 
In-situ sterilizable and autoclavable
  • Designed for operation under fully aseptic conditions and to withstand repeated sterilization at 130 °C (266 °F) without any decline in measuring reliability.
 
Long operational electrode lifetime
  • The electrode is provided with a patented silver-ion trap which prevents silver sulfide precipitation at the diaphragm in sulfide-containing media and extends electrode life.

الوثائق

Certificates

Confirmation Specifications Silicone O-Rings 10.77x2.62 for pH-Electrodes
Specifications confirmation 203011136IGConfirmation for the following electrodes:InPro3030, InPro31XX, InPro32XX, InPro3300, InPro4010, InPro42XX, InP...
الحصول على عرض الأسعار
Pt4805-DPAS-SC-K8S Redox/ORP Probe

الملحقات

الحصول على عرض الأسعار
Pt4805-DPAS-SC-K8S Redox/ORP Probe
Thank you for visiting www.mt.com. We have tried to optimize your experience while on the site, but we noticed that you are using an older version of a web browser. We would like to let you know that some features on the site may not be available or may not work as nicely as they would on a newer browser version. If you would like to take full advantage of the site, please update your web browser to help improve your experience while browsing www.mt.com.