Pipette Calibration and Pipette Testing
Pipette Calibration

Pipette Check & Calibration

Individual Solutions from Pipette Testing to Calibration Services

 

Pipettes Calibration & Checking FAQ

Below you will find answers to frequently asked questions related to Pipettes Check & Calibration and Good Pipetting Practice.


Below you will find answers to frequently asked questions related to Pipettes Check & Calibration and Good Pipetting Practice.


 

What is a pipette?

A pipette is a laboratory tool to transport small amounts of measured liquid accurately by aspiration and dispensing the sample. Today the majority of pipettes are piston pipettes and create a partial vacuum above the liquid to aspire the liquid, herewith avoiding contacting mechanics.

More: Rainin Pipettes

What is a calibrated pipette?

A pipette is calibrated when the expected volume with the actual volume delivered by a pipette are determined together with the associated measurement uncertainty by an accredited institution. Purpose is to prove if the pipette is still within its manufacture or ISO8655 specifications to deliver accurate results. Pipettes out of tolerance can often be serviced, adjusted and re-calibrated.
More: Rainin Pipette Service

Where is a pipette used?

Piston Pipettes are typically used in chemistry, biology, clinical and medical areas, where liquid samples are taken and must be transported often in small and accurate quantities for further dilutions, research and analysis work.

How to use a pipette?

Here are some very common pipetting techniques and errors for piston pipettes and how to avoid them.

  1. Wrong tip immersion depth
    The correct tip immersion depth can improve accuracy by up to 5%. The tip should be immersed between 1-2 mm for micro-volume pipettes and up to 3-6 mm for normal-range pipettes, depending on the tip size. If the tip is immersed too far, the volume of gas in the tip is compressed, causing too much liquid to be aspirated.

  2. Wrong pipetting angle
    The immersion angle of your pipette tip in the sample should be as near to vertical as possible and should not deviate more than 20 degrees from vertical. A more horizontal angle causes too much liquid to be drawn into the tip, resulting in inaccurate aspiration. For example, at an angle of 30 degrees to vertical, up to 0.7% too much liquid can be aspirated.

  3. Inconsistent dispensing
    Greater accuracy and sample-to-sample reproducibility can be achieved by ensuring that the last remaining droplet dispenses fully and does not adhere to the tip end. For most applications, dispensing with the end of the tip resting against the vessel wall is recommended as it reduces or eliminates the amount of sample remaining in the tip. This technique can increase accuracy by 1% or more.

  4. Failure to pre-rinse
    Dispensing liquid from a pipette leaves a coating of the liquid on the tip, making the expelled volume slightly less than it should be. Pre-rinsing a new tip at least twice with the liquid about to be used will condition the inside of the tip.

  5. Inconsistent pipetting rhythm
    Use a consistent pipetting rhythm from sample to sample. Avoid hurrying or rapid operation and get into a rhythm for each step in the pipetting cycle.

    More: Good Pipetting Practice

How to check pipette accuracy?

Pipette accuracy shall be checked regulary between calibrations, to avoid out of tolerance results between calibrations. To check the pipette at least the nominal volume shall be dispensed in a beaker on an accurate analytical balance at least 4 - 6 times of which the average volume and deviation can be calculated using ISO8655 formulas. Balances with a built-in application or PC software deliver a Pass or Fail result within seconds automatically.

More: Pipette Check

How to check pipette calibration?

Pipette accuracy shall be checked regulary between calibrations, to avoid out of tolerance results between calibrations. To check the pipette at least the nominal volume shall be dispensed in a beaker on an accurate analytical balance at least 4 – 6 times of which the average volume and deviation can be calculated using ISO8655 formulas. Balances with a built-in application or PC software deliver a Pass or Fail result within seconds automatically.  

Why is pipette calibration important?

A pipette calibration shows the difference between expected and actual volume, which shall be within a given tolerance. If it is outside tolerance it can produce falsch dilutions and in the end even wrong analytical results. As a consequence results can't be reproduced and whole analysis have to be redone. To detect an out of tolerance pipette early, it is recommended to check the pipette accuracy also between calibrations to avoid that the out of tolerance is first detected at the next calibration.

What is pipette calibration?

With a pipette calibration the expected volume with the actual volume delivered by a pipette are determined together with the associated measurement uncertainty by an accredited institution. Purpose is to prove if the pipette is still within its manufacture or ISO8655 specifications to deliver accurate results. Pipettes out of tolerance can often be serviced, adjusted and re-calibrated.

How to clean a pipette?

Before start cleaning first always consult the manufacture manual, because material resistance differ per brand. For a good cleaning, piston pipettes need to be dissembled. The tip ejector arm and cylinder can often simply unscrewed by hand to open the pipette. Be careful: the spring pressure can cause that parts jump off. Parts can typically be cleaned with a cloth lightly soaked with a mild soap with distilled water or using an Isopropyl alcohol or Ethanol solution. Avoid that liquids can enter the shaft, also during pipetting to avoid  contamination of the piston and seal. For sterilization or autoclaving, look into the manual if and how this can be done for the desired pipette model. After cleaning the pipette shall be check if it is within tolerance. Typically a pipette service technician performs a calibration after service to verify and confirm that the pipette is within tolerance.

Pipette Check Solution
Pipette Calibration Procedure

Pipette Check

Pipette Calibration

Solutions that enable Quick Check within minutes up to daily decicated performance checks supported by RFID and software that track the status of your pipette anytime anywhere.
METTLER TOLEDO's weighing reputation combined with process oriented software result in efficient, reliable and compliant pipette calibration process according ISO8655.
Technical Specifications
Capacity up to 220 g and Readability to 0.01mg
Pipette check range
20 µl to 10 ml
Evaporation Trap
With SmartCheck your avoid evaporation for better results
Guided Operating
For fast and consistent pipette checks
GPP - Good Pipette Practice
Regulary checks according SOP are done quickly within minutes
RIFD Pipette tracking
For faster and reliable checking of pipettes
Technical Specifications
Capacity up to 220g and Readability up to 1µg
Pipette Calibration range
1 µl to 10 ml
Motion sensor trap lid
For best reproducible results even with micro pipettes
Process Operation
Professional and with audit trial using Calibry software
Regulary Compliance
Fulfills ISO8655 or other defined custom methodes
Reporting
Create traceable compliant reports with Calibry software
Pipette Check Solution

Solutions that enable Quick Check within minutes up to daily decicated performance checks supported by RFID and software that track the status of your pipette anytime anywhere.

Technical Specifications
Capacity up to 220 g and Readability to 0.01mg
Pipette check range
20 µl to 10 ml
Evaporation Trap
With SmartCheck your avoid evaporation for better results
Guided Operating
For fast and consistent pipette checks
GPP - Good Pipette Practice
Regulary checks according SOP are done quickly within minutes
RIFD Pipette tracking
For faster and reliable checking of pipettes
Pipette Calibration Procedure

METTLER TOLEDO's weighing reputation combined with process oriented software result in efficient, reliable and compliant pipette calibration process according ISO8655.

Technical Specifications
Capacity up to 220g and Readability up to 1µg
Pipette Calibration range
1 µl to 10 ml
Motion sensor trap lid
For best reproducible results even with micro pipettes
Process Operation
Professional and with audit trial using Calibry software
Regulary Compliance
Fulfills ISO8655 or other defined custom methodes
Reporting
Create traceable compliant reports with Calibry software

Services

(Toll Free) 1800 22 8884 & 1800 1028 460
Call Service

Explore our Services - Tailored to Fit your Equipment

We support and service your measurement equipment through its entire life-cycle, from installation to preventive maintenance and calibration to equipment repair.

Uptime
Support & Repair
Performance
Maintenance & Optimization
Expertise
Training & Consulting
 
 
 
 
 
 
 
Thank you for visiting www.mt.com. We have tried to optimize your experience while on the site, but we noticed that you are using an older version of a web browser. We would like to let you know that some features on the site may not be available or may not work as nicely as they would on a newer browser version. If you would like to take full advantage of the site, please update your web browser to help improve your experience while browsing www.mt.com.