Need assistance?
Our team is here to achieve your goals! Speak with our experts.

DSC Calibration, Heating Rate Independence

Sample

Zinc (calibration standard, purity > 99.999 %)

 

Application

Standard for calibration

 

Conditions

Measuring cellDSC821e

Pan: Alumina 40 µl, with pierced lid

Sample preparation: Zinc pellet, pressed flat and premelted

DSC measurement: Heating from 350 °C to 475 °C at 5, 10 and 20 K/min. All measurements are performed with the same sample. Cooling rate is 5 K/min.

Atmosphere: Nitrogen, 50 cm3/min

 

 

Interpretation

The DSC curves show the melting of zinc at different heating rates. If displayed with respect to temperature, the peak area increases with increasing heating rates. The heat of fusion is however the integral of the heat flow with respect to time. This is just as independent of the heating rate as the melting point.

 

Evaluation

The onset temperature and the heat of fusion of zinc are determined.

 MeasuredMeasuredMeasuredRef. value
Heating rate51020K/min
Melting point (onset)419.5419.6419.8419.7 °C
Heat of fusion107.0107.0107.1107.0 J/g

 

Conclusion

The calibration data of an instrument are affected by many factors such as the heating rate, the purge gas used, the sample pan material and the temperature range used. Only when these effects can be taken into account in the calibration data, are results obtained that are independent of measurement parameters, as is shown in this example of the melting point and the heat of fusion.

 

DSC Calibration, Heating Rate Independence | Thermal Analysis Application No. HB804 | Application published in METTLER TOLEDO TA Application Handbook Pharmaceuticals