PP from Different Manufacturers - METTLER TOLEDO

PP from Different Manufacturers

Sample

Ferrolene HU10TS3, Hostalen PPH1050, Hostalen PPR 1042, Hostalen PPX 4305, Procom GC30H251, Vestolen P7032G, all as pellets

 

Conditions

Measuring Cell: DSC20

Pan: Aluminum standard 40µl, pierced lid

Sample Preparation: Disk cut from center part of pellet

DSC measurement: Heating from 30 to 250ºC at 10 K/min

Atmosphere: Nitrogen, 50 cm3/min




Conclusion

Polypropylene can easily be characterized by DSC.

 

PP from Different Manufacturers | Thermal Analysis Handbook No.HB214 | Application published in METTLER TOLEDO TA Application Handbook Thermoplastics