Watch our videos on how to get started with EasyPlus UV/VIS spectrophotometers. Learn how to perform different measurement methods, how to quickly replace the lamp of an Easy VIS spectrophotometer, and much more!
EasyPlus UV/VIS spectrophotometers combine conventional spectrophotometry, color measurements, and water analyses within one compact instrument. The spectrophotometers are designed to facilitate your routine spectroscopic workflows, saving bench space and reducing your investment in equipment and service.
Getting Started
Unboxing Your EasyPlus UV/VIS Spectrophotometer
It is easy to get started with the new EasyPlus Spectrophotometers by METTLER TOLEDO. Simply:
- Unpack,
- Connect,
- Adjust,
and perform your first measurement in less than 5 minutes!
First open the package and remove the following parts. User manual. Box of a pair of standard level cuvettes. Power supply with cable. and the dust cover. Lift up the instrument and remove the two large foam parts. Place the instrument on the bench and remove the stickers and the foam below the front cover. Close the lid. Now insert the plug of the power cable into the socket of the AC-DC adapter. Insert the plug of the AC-DC adapter into the 24-volt socket on the rear side of the instrument. Tighten the knurled nut to secure the plug. Plug in the power cable into a grounded power outlet which is easily accessible. Press the power button to switch on the instrument. Select a language and check by opening the lid whether the sample compartment is empty. The shown pop-up will be confirmed by closing the lid. Wait until the initialization is finished. After the first startup, you have to perform adjustments. Go to Setup and Tools, Adjustments and select it. System baseline and dark current for EasyVis and system baseline for EasyUV. Now your instrument is ready to use.
METRA Toledo's Easy Plus UV-VIS spectrophotometers are supporting spectroquant test kits. But there is no button on the screen? You can order separately the SQ kit reader from METRA Toledo. In this video you will learn how to connect it easily. First switch off the instrument and unplug the power cable. Take the Allen key from the storage compartment below the back cover. Open the lid and remove the shown cover. Mount the SQ Kit reader on the magnetic pogo pins. Use the delivered screw to fix the SQ Kit reader. Tighten the screw completely to ensure the correct positioning. Now Plug in the power again and switch on the instrument. After initialization, the spectroquant button appears on the home screen. Click spectroquant. Open the lid and scan the barcode of the test kit. The method starts automatically. By closing the lid, the measurement starts immediately. For reagent test kits, the barcode on the outer selector can be scanned. The instrument checks the allowed path length and is informing which ones are allowed. By moving the X-path holder into the correct position, the pop-up will disappear. If a result is out of limit, L_O_ or H_I_ together with an arrow is shown. To perform the next analysis, click the start button and follow the instructions on the screen.
Installing the SQKitReader Accessory For Spectroquant Water Test Kits
Learn how to easily install the SQKit reader accessory for Spectroquant water test kits on your EasyPlus UV/VIS spectrophotometer from METTLER TOLEDO.
Simply mount the SQKitReader on the magnetic pogo pins of the EasyPlus UV/VIS spectrophotometer and fix it with the included screw. After the initialization, the Spectroquant button appears on the home screen, enabling over 150 water test methods.
Internal Barcode Reader Set SQKitReader
More about Photometric Water Testing Applications
How to Measure
How to Perform Photometric Measurements with EasyPlus UV/VIS Spectrophotometers
In this video, you will learn how to easily perform photometric measurements on your EasyPlus UV/VIS Spectrophotometer from METTLER TOLEDO.
You can easily configure and save your own methods directly to the instrument. Visual guidance on the terminal walks you through the different measurement steps.
In this video I will show you how to perform photometric measurements with Metlar Toledo's easy plus UV-VIS spectrophotometers. The photometric methods determines the absorbance and or transmittance at up to five different wavelengths. I will show you how to easily configure a method and perform the measurement afterwards. Let's start with the configuration. Click Photometric on the home screen. This opens a list of 20 predefined photometric methods. Each method can be adjusted individually and exported to an USB flash drive. To enter the method editor, just click on the edit button and select your method. The method editor shows two tabs. General and Measurement The General tab is the same for all measurement categories. You can change parameters, for example the method name, you can choose between different export options or define sample IDs. The measurement tab contains all measurement information. You can define the path length. If you're interested in absorbance, transmitters or both, you can also select the number of wavelengths, five at maximum. If selecting variable wavelength I decide whether I want to define a fixed wavelength now, or to be prompted before each measurement. For now I will define here a wavelength at 690 nanometers. If I activate the background correction I need to define the corresponding wavelength. The default value is 840 nanometers. Last but not least, I can also define the number of decimal places per result, meaning absorbance and or transmittance. Now the method is completely defined and we are ready to perform the measurement. To start the method I'm clicking the green start button. The instrument now guides me through the whole workflow. At first I need to perform the blank measurement. I follow the instruction shown on the screen and press the lid to open. I'm inserting the blank cuvette, close the lid and the blank determination starts automatically. Now I'm asked to insert the sample. Let's open that lid again and replace the blank with the sample cuvette. Close the lid and the measurement starts immediately. And here is the result. To measure additional samples with the same settings, simply click the start button again. I am now directly asked to insert the sample. That's because the blank measurement is stored for seven days and no new blank determination is necessary, even if I am leaving the method. So just simply follow the instruction to continue with further measurements. I hope this short introduction into photometric measurements was helpful. Don't miss our other. training videos on scanning, quantification, color and water testing methods. See you soon.
In this video, I will show you how to perform a scan measurement with Merta Toledo's EasyPlus UV-Vis spectrophotometers. With a scan method, a full spectrum in absorbance or transmittance can be recorded. Peaks and valleys can be detected automatically. I will show you how to easily configure a method and perform the measurement. Let's start with the configuration. Click Scan on the home screen. This opens a list of 20 predefined scan methods. Each method can be adjusted individually and exported to an USB flash drive. To enter the method editor just click on the Edit button and select your method. The method editor shows two tabs General and Measurement. The general tab is the same for all measurement categories. You can change parameters, for example the method name. You can choose between different export options or define sample IDs. The Measurements tab contains all measurement information. You can define the power flange, the wave length range, for example 400 nanometers to 800 nanometers, the scan speed between low, medium and high. I will go for medium. The step size. 0.5, 1 nanometer, 2 or 5 nanometers, I will go for 5 nanometers. The result of interest? absorbance or transmittance. And last but not least, I can activate the peak valley detection and can adjust the sensitivity and the wavelength range. Now, the method is completely defined and we are ready to perform the measurement. To start the method I'm clicking the green start button. The instrument now guides me through the whole workflow. At first, I need to perform the blank measurement. I follow the instruction shown on the screen and press the lid to open. I'm inserting the blank cuvette, close the lid and the blank determination starts automatically. The scanning progress is shown with this moving bar. Now I'm asked to insert the sample. Let's open that lid again and replace the blank with the sample cuvette. I close the lid and the measurements starts immediately. The spectrum is shown live on the screen. The detected peaks and valleys are labeled in the spectrum but can also be found in the tab Peaks and Valleys. To measure additional samples with same settings Simply click on the start button again. I'm now directly asked to insert the sample. That's because the blank measurement is stored for seven days and no new blank determination is necessary, even if I'm leaving the method again. Simply follow the instructions to continue with further measurements. I hope this short introduction into scan measurements was helpful. Don't miss our other product training videos on photometric and color measurements, quantification and water testing. See you soon!
How to Perform Scan Measurements with EasyPlus UV/VIS Spectrophotometers
Take a look at how to easy it is to perform scan measurements on your EasyPlus UV/VIS Spectrophotometer by METTLER TOLEDO.
With the scan method, a full spectrum absorbance or transmittance can be recorded. Peaks and valleys can be detected automatically by the instrument.
How to Perform Quantification with EasyPlus UV/VIS Spectrophotometers
Performing quantification measurements on your EasyPlus UV/VIS Spectrophotometer by METTLER TOLEDO is simple. They can be performed using two different types of result calculations:
- By an own formula
- Or with a calibration curve, built up by measuring standards
The configuration of both methods is described in detail within this video.
In this video I will show you how to perform quantification measurements with the Mettler Toledo's easy plus UV -Vis spectrophotometers. Quantification contains two different types of result calculation via an own formula or by calibration curve built up by measuring standards. I will show you how to perform both different types of quantification measurements. Let's start with the configuration. Click Quantification on the home screen. This opens a list of 20 predefined quantification methods. Each method can be adjusted individually and can be exported to a USB flash drive. To enter the method editor, just click on the Edit button and select your method. The method editor shows two tabs, General and Measurement. The General tab is the same for all categories. You can change for example the method name into blue, you can choose between different export options or define sample IDs. The Measurements tab is different for formula and calibration curves. Firstly, I will start with explanation of formula. If you want, you can also jump directly to calibration curve. The measurement tab contains all measurement information. First of all, we have to choose the calculation type. I'm now selecting formula. Additionally, you can define the power of length. For example fixed and 60 mm. Here up to five measurement wave lengths can be defined. The measured absorbance and transmittance values can be used later in the calculation of the result. I will define 820 nm as a wavelength. With constants you can define up to ten constants which can be used in the calculation afterwards. For the first constant I'm defining 5.5. Sample factors are flexible constants which are regressed before each measurement. You can define up to three of them and define also the name and the default value. For example, X'name is weight. Last but not least, I can define the results where the formula can be input. This list can contain up to four results. For each result, name, unit, formula, limits and decimal places can be defined. The formula editor follows a building block logic. You have four buttons selecting your building blocks. The first one contains the measured absorbance and transmittance values. The second one the previously defined constants. The third one offers mathematical operators. And the last one adds sample factors and brackets to your formula. By confirming the input, the formula is ready to use. Now the method is completely defined and we are ready to perform the measurement. To start the method I'm clicking the green start button. The instrument now leads me through the whole workflow. First of all the instrument is requesting the sample factor which I define with 0.5. Afterwards I need to perform the blank measurement. I follow the instruction shown on the screen and press the lid to open. I'm inserting the blank cuvette, close the lid, and the blank determination starts automatically. Now I'm asked to insert the sample. Let's open the lid again and replace the blank with the sample cuvette. I close the lid. And the measurement starts immediately. And here is the result. To measure additional samples in the same settings, simply click the start button again. I'm now again asked for the sample factor and by confirming the input the measurement starts automatically. That's because The blank measurement is stored for seven days and no new blank determination is necessary, even if I am leaving the method. Simply follow the instruction, continue with further measurements. Let's have a look how to set up a calibration curve. The measurement tab contains all measurement information. First of all, you have to choose the calculation type. I'm selecting Calibration Curve. Additionally, you have to define the path length. If I activate the background correction, I need to define the corresponding wavelength. The default value is 849 nanometers. A name and a unit can also be defined here. For example, yellow is the name. On the second page I can define the decimal places. Here you have to define the wavelength of interest, for example 340 nanometers. For the curve different fit types are available, linear, quadratic or cubic. For a linear fit I can also activate the setting Force through zero. In the next step I have to define the standard for the calibration curve. By clicking the Add button I can define the first data point with the corresponding concentration, five, and press afterwards the Start button. If no blank value is available I have to measure the blank value first. Afterwards The instrument is re-casting for a standard solution. The result is shown and I'm confirming it. By returning to the list of data points, I can define more data points. I finished with the data point measurements. I can check the calibration curve formula here. Here we can see the formula and the A square value. Now, the method is completely defined and we are ready to perform the measurement. To start the method, I am clicking the green start button again. The instrument now leads me through the whole workflow. Now I'm asked to insert the sample. Let's open the lid again and I'm inserting the sample. I close the lid and the measurement starts immediately. And here is the result. To measure additional samples at the same settings, simply click the Start button again. I'm now directly asked to insert the sample. That's because the blank measurement is stored for seven days and no new blanket elimination is necessary, even if I'm leaving the method. Simply follow the instruction to continue with further measurements. I hope this short introduction into quantification measurements was helpful. Don't miss our other product training videos on photometric and scanning measurements. Color determinations and water testing. See you soon.
In this video, I will show you how to perform a color measurement on Mettler Toledo's EasyPlus UV -Vis spectrophotometers. The Easy UV instrument has 30 predefined color calculations. I will show you how easily to configure an own method and perform a measurement. Let's start with the configuration. Click Color on the home screen. This opens a list of 20 predefined color methods. Each method can be adjusted individually and exported to a USB flash drive. To enter the method editor, just click on the Edit button and select your method. The method editor shows two tabs. General and Measurement The general tab is the same for all measurement categories. You can change parameters, for example the method name. You can choose between different export options or define sample IDs. The measurement tab contains all measurement information. You can define the path length, choose a color scale from the list, for example CIELAB color. You can choose between different illuminance and observers, you can define the step size, 1 or 5 nanometers. To follow the official calculation with 10 millimeter path lengths, you can use normalization. With this parameter activated, the measured transmittance values are automatically recalculated to this path length. On the second page, you can select the wavelength range, for example 360 to 830 nanometers. And, last but not least, I can define my own result limits. Now the method is completely defined and we are ready to perform the measurement. To start the method I'm clicking the green Start button. The instrument now guides me through the whole workflow. At first I need to perform the blank measurement. I follow the instruction shown on the screen and press the lid to open. I'm inserting the blank cuvette, close the lid and the blank determination starts automatically. The scanning progress is shown with this moving bar. Now I'm asked to insert the sample. Let's open the lid again, replace the blank cuvette with the sample cuvette, close the lid and the measurement starts immediately again. The spectrum is shown live on the screen. After the completed measurement, the color number is immediately shown. Due to a flexible integration time, the measurement takes longer. when the transmittance is very low. This feature ensures better repeatability and better accuracy of the measured spectrum. The results are displayed all in green because result limits were activated and all results are within the defined limits. To measure additional samples with the same settings, simply click the Start button again. I'm now directly asked to insert the sample. That's because the blank measurement is stored for seven days and no new blank determination is necessary, even if I'm leaving the method. Simply follow the instruction to continue with further measurements. I hope this short introduction into color measurements was helpful. Don't miss our other product training videos on photometric and scanning measurements, quantification and water testing. See you soon.
How to Perform Color Measurements with EasyPlus UV/VIS Spectrophotometers
The Easy UV offers 30 predefined color calculations in the UV and visible wavelength range. For applications only measuring in the visible region, the Easy VIS features 25 color numbers.
In this video, you will learn how to easily perform color measurements on your EasyPlus UV/VIS Spectrophotometer from METTLER TOLEDO.
Configuring a method is very simple, and each method can be adjusted individually and exported to a USB flash drive. Via the normalization parameter, all measured transmittance values can be automatically recalculated to a 10 mm path length.
How to Perform Water Testing with EasyPlus UV/VIS Spectrophotometers
In this video, you will learn how to easily perform water testing measurements with Spectroquant™ test kits on your EasyPlus UV/VIS Spectrophotometer by METTLER TOLEDO.
In this category, more than 150 predefined methods are available. Scanning the 2D barcode of a test kit with the SQKitReader™ accessory automatically starts the corresponding method.
In this video, I will show you how to perform the spectral quant method with Mettler Toledo's EasyPlus UV-Vis spectrophotometers. In this category, more than 150 pre-defined methods are available. By scanning the 2D barcode of a test kit, the corresponding method starts automatically. This ensures the correctness of the results. Let's start with the first test kit. Click spectro quant on the home screen. This activates the SQ kit reader. To enter the method I have to scan the barcode on the test tube. I have to open the lid and turn the cuvette carousel into the 60 mm position. Now I can insert the tube. I have to ensure that the barcode is aligned with the SQ kit reader. During inserting the tube, the instrument starts automatically the correct method. If no zero adjustment is available, I am requested to insert a zero cell. I am replacing the sample tube with a zero cell and close the lid to start the measurement. To measure a used aeration blank, I click on Blanks and tap on the slider. As no used aeration blank value is stored, I'm confirming to measure one now. The instrument asks me to insert the aeration blank cell. I'm inserting the aeration blank cell, close the lid and the measurement starts automatically. By clicking the Start button I can repeat the measurement and the instrument is calculating the mean value of the replicates. By clicking the Back button I am confirming the reagent blank measurements. The letter R indicates the activated user reagent blank. In settings tab, I can define the desired citation, the desired unit and other test kit specific parameters. Now the method is ready to use. By clicking on sample, I'm requested to insert the sample. I'm inserting the sample cell, close the lid and the measurement starts automatically. To measure additional samples with the same settings, Simply click the Start button again. I now directly ask to insert the sample. On the Result screen, I can change the citation and unit, and the result will be recalculated automatically. For test kits with an autosampler, the barcode can easily be scanned. Afterwards, the instrument is informing which path lengths are allowed for this test kit. The pop-up can be confirmed by moving the X-Path holder into the correct position. The instrument asked me to insert the zero cell again. Close the lid and the zero adjustment measurement starts automatically. Now I'm asked to insert the sample. Let's open the lid again, replace the zero cell with a sample cuvette, I close the lid and the measurement starts immediately. And here is the result. To measure additional samples with the same settings, simply click the start button again. I'm now directly asked to insert the sample. That's because the blank measurement is stored for seven days. and no new blank determinations necessary – even if I'm leaving the method. I hope this short introduction into spectroquant measurements was helpful. Don't miss our other product training videos on photometric, scanning, color and quantification measurements. See you soon.
Welcome, I'm our Mettler Toledo application lab here in Switzerland. In this video I will explain how you can enable Mettler Toledo's EasyPlus UV-Vis spectrophotometers measuring nanocolor water test kits from Macherey -Nagel. The video will also demonstrate where the methods can be downloaded and how the handheld barcode reader can be used to facilitate your daily work. To illustrate the workflow I will use a COD Nano Color test kit. To get started, we first need to download a method database from mc .com. The link is provided below the video. Once the database for either the easyVIS or easyUV spectrophotometer has been successfully downloaded, the folder must be unzipped and stored in the root directory of a USB flash drive. All methods for Nanocolloid test kits are quantification methods. Therefore, you can import your desired methods into this quantification category. To do so, connect the flash drive to the instrument and navigate to the method list where you will find an import-export button. Select the method you want to overwrite and you will be presented with a list of all available methods on the USB flash drive. Select the one that corresponds to your test kit. In this example I will import the COD method 985-038. In order to facilitate the workflow, each method comes with a predefined barcode for the ScanStart functionality. Scan start enables to start a method through the scanning of a barcode instead of selecting the method from the list. Furthermore, the barcode reader can be utilized for the input of the sample ID. But where do you find the barcodes? In the same method zip file you just downloaded, you will find a PDF template to print the barcodes for your testkits. I will select here the COD test kit 985038 from the drop-down list. The layout of this document is compatible with Avery labels in D-N -A4 and U.S. letter format as shown here. Let's proceed to the printer and pick up the label. Having now printed our barcode on the label sheet, we are ready and prepared to analyze our sample. So let's go to the lab and start the workflow. I am fixing the barcode label to the lid of the test kit box, but you could also put it onto your SOP. Each Nanocolor test kit includes a description of the sample preparation process in the lid of the box. By following the instructions you can be confident that you will obtain accurate results. I have prepared the sample and the digestion step in the thermoreactor is already complete. By scanning the barcode, the method starts automatically. You will then be guided step by step through the measurement procedure. First insert the blank and close the lid. Once the blank measurement is complete You will be prompted to insert the sample. The sample is displayed on the screen. I hope this video was useful in helping you to understand the workflow for NanoColor test kits. Additionally, I would like to mention that individual barcodes can also be generated for all your customized methods. To do so, just add the barcode information in the parameter method start in the method editor tab general. Thank you for watching this video and don't miss our other videos explaining the different measurement categories. Scanning. photometric quantification and color measurements.
How to Run NANOCOLOR Water Test Kits on EasyPlus UV/VIS Spectrophotometers
In this video, we explain how you can enable METTLER TOLEDO's EasyPlus UV/VIS spectrophotometers measuring NANOCOLOR® water test kits from Macherey-Nagel. This video will also demonstrate where the methods can be downloaded and how the handheld barcode reader can be used to facilitate your daily work.
How to Use the EasyDirect™ UV/VIS Software
Learn how to take full advantage of EasyPlus UV/VIS spectrophotometers in this informative video. Connecting your instrument with the user-friendly EasyDirect UV/VIS PC software allows you to streamline your daily workflow and simplify your analysis process.
Join us as we demonstrate the benefits of using this software and guide you through the easy connection process. Make the most of your EasyPlus spectrophotometer with the EasyDirect UV/VIS PC software.
Welcome for my application lab here in Switzerland. In this video I will show you how you can unlock the full potential of your easy plus UV-Vis spectrophotometers by connecting them to the easy direct UV-Vis PC software. You will learn how the software can make your daily work easier and how simple it is to connect the instruments to this software. Let's start with a quick introduction to the easy Direct UV-Vis software itself. We have three main views: Instruments, Measurement Results and Methods. Below Instruments you can see a list of connected instruments, their firmware status and the results executed performance test methods. If the firmware of your instrument is outdated, you can download the latest version via EasyDirect to a USB flash drive. In this instrument view, you also have the possibility to define an automatic export from EasyDirect to a storage location of your choice. This can be done in certain time intervals or every time when a new result is transmitted to the software. The feature enables the automatic export for example to a folder managed by your local link system. Below measurement results you can find all results which were transmitted to the database. We will have a closer look to this view later. Below methods you can store and restore backups of your connected instruments. With this feature you only have to define the methods once on an instrument. Afterwards, you can restore the corresponding backup also on other EasyPlus UV-Vis spectrophotometers. As mentioned before, you will now have a closer look into the view measurement results. On the top you can select for which instrument you want to display the results. Here, on the left hand side, you find the same measurement categories as on the instrument. Each category looks similar: you have the big results table in the middle and additional panels on the right-hand side displaying result entry specific data. The layout of this main table can be configured and saved via the button column layout. This needs to be done for each category separately. You can also apply filter to the main table. The date filter is located here on top. But you can also filter in each column by clicking on the funnel symbol in the column header. So much for the theory. Let's have a look at the software live in action. To add an instrument to your EasyDirect UV-Vis software, you just need to plug in a USB/A -B cable from the instrument to your computer. Software will automatically detect the new device and is asking you whether you want to edit. METRA ToLido is offering two different license types, one for the connection of a single instrument and one for the parallel connection of up to five instruments. After establishing the connection you have to activate the parameter export to EasyDirect in your method on the instrument. This parameter can be found in all categories in the general tab. Per default the export is enabled. As soon as you execute a method, the result will be transmitted automatically into the easy direct database. With a simple scan measurement I want to show you how quick the transmission is and what you can do with the data afterwards. Here I have prepared a scan method with the enabled method parameter export to easy direct. As soon as the spectrum scan is finished, we can find new entry in the result list of the easy direct UV -Vis PC software. If we select the measurement category Scan, we can find our result and on the right hand side a list of peaks and valleys, the spectrum itself and some additional information. To analyze the spectrum in more detail, we can choose between two options. Graph and spectra graphics. With graph the spectrum will be displayed and you can zoom in and stack spectra with a multiple selection of entries. With spectra graph the spectral data will be sent to an additional external software. With this spectroscopy software you can do additional evaluations like peak finding or calculating derivatives. If you are regularly measuring color samples according to different color scales, the EasyDirect UV-Vis software can save you precious time as it can show multiple color results out of a single measurement scan. Let me quickly demonstrate this to you. I have here the yellowish sample from which I want to determine the color according to the Zeebold and ASTM color scale. I initiate the measurement by starting the prepared Zeebold method on the instrument. And you can see the spectrum on the screen. As soon as the measurement is finished, the color result is displayed and transferred to the Easy-DIRECT PC software. Here we can see now the measured spectrum and the corresponding Zewalt color result. With a button color scales I can now define two additional color scales depending on the main result from the instrument. In our case I have measured zebra color with the instrument and EasyDirect will directly show us the corresponding ASTM color. So let's directly edit. And as you can see the result of AST color now also is a part of the database and can be exported. This saved us a whole measurement cycle. I hope this video was helpful to understand the key functionalities of our easy direct UV-Vis PC software. Thank you for watching this video and please also look into our other how-to videos for easy plus UV-Vis spectrophotometers explaining the various measurement categories, photometric, scan, color measurements and water testing. See you there.
Easy Maintenance
How to Replace the Lamp of Your Easy VIS Spectrophotometer
METTLER TOLEDO's Easy VIS Spectrophotometer operates within a wavelength range of 330 to 1,000 nm. The light source is an easy-to-replace tungsten lamp, which any operator can quickly replace within just a few minutes! Simply unscrew the screws with the supplied tool, replace the module, and close it again.
Easy VIS constantly monitors and records the usage of the lamp and prompts the user to prepare for a replacement in advance. If the Easy VIS detects that the light intensity is below a threshold, it informs the user.
In this video, you will learn how easily the lamp can be replaced on the EasyViz spectrophotometer from Mettler Toledo. When the lamp has reached its lifetime, this pop-up is shown to you. Switch off the instrument and wait for at least 30 minutes to cool down the lamp module. Remove the back cover and take the Allen wrench from its position. Unplug the power cable and tilt the instrument on its right side. Open the cover. Disconnect the cable by pressing down the latch and pulling the plug out of the socket. Unscrew the two screws with the Allen wrench. Pull out the lamp module. Unpack the new lamp module. Do not touch the light bulb. Insert the new lamp module and tighten the two screws with the Allen wrench. Insert the lamp plug into the lamp socket and close the cover. Return the instrument to its upright position and plug in the power cable. Switch on the instrument. After initialization, go to Setup and Tools and reset the lamp counter by clicking Lamp Reset. After a warm-up time of at least one hour, you have to perform an adjustment. Go to Setup and Tools, Adjustments and select System Baseline. Press the start button and confirm that the compartment is empty. Now your instrument is again ready to use.