Webinar: Pharmacopeia – What to Check on UV/Vis Spectrophotometers?
During the webinar, we will go into depth about these questions and more:
What certified reference materials (CRMs) are specified for wavelength accuracy and precision tests by both Pharmacopeias?
USP: Rare earth oxide solutions/glasses (Ce, Ho, Didymium) or Hg/D2 emission lines
Ph. Eur.: Ho(ClO4)3, Ce2(SO4)3 solutions, or Ho/Didymium glasses, or Hg, D2, Ne, Xe lines
What CRMs are specified for photometric accuracy and precision tests by both Pharmacopeias?
USP: for UV range, Nicotinic acid in HCl (6 – 60 mg/L, up to 2.5 A, 213 and 261 nm) or K2Cr2O7 in HClO4 (6 – 200 mg/L, up to 3.0 A, 235, 257, 313, 350 nm); for visible range, neutral density glass filters (up to 3.5 A, 440, 465, 546.1, 590, and 635 nm)
Ph. Eur.: Suitable solid and liquid filters such as certified PDC or neutral density filters; or, instead of certified PDC, Nicotinic acid in HCl.
What CRMs are specified for photometric linearity tests by both Pharmacopeias?
USP and Ph. Eur., respectively, specify to use the materials for photometric accuracy tests also for photometric linearity tests.
Lab Managers, Lab Technicians, System Owners, Quality Assurance Managers, Quality Assurance Operators

Dr. Fabian Müller
Product Manager SPG UV/VIS
Dr. Fabian Müller, Ph.D., is an accomplished chemist from the ETH Zürich in Switzerland. After completing his doctoral studies, Dr. Müller joined the team at METTLER TOLEDO as an Application Specialist for UV/Vis spectroscopy, where he quickly established himself as a skilled and knowledgeable professional. His expertise in this area led to his promotion to the position of Product Manager for UV/Vis spectrophotometers.