What factors cause interference in dissolved ozone measurements?

How long does the initial setup of an ozone sensor take before I can get an ozone reading?

My ozone sensor readings have become unstable or drifted significantly. What should I check?

For electrochemical ozone sensors, you should work through the most common causes based on the symptom you are seeing:

Ozone Frequently Asked Questions

Unstable zero current

Check that the membrane surface is completely dry during zero calibration, then check for air bubbles trapped in the electrolyte. If the electrolyte is depleted, replace the full membrane and electrolyte assembly, reassemble, and then recalibrate.

Slope trending in one direction over several calibrations

 This indicates that the electrolyte needs to be replaced.

Slope too low

Check for cathode contamination. Clean the cathode gently with warm, soapy water and rinse with deionized water.

Slope too high

This means the membrane could be expired and should be replaced.

What dissolved ozone concentration is required to sanitize a pharmaceutical WFI loop?

How often do dissolved ozone sensors require calibration?

How much maintenance is required for an ozone sensor, and how can I avoid significant downtime?

How does Intelligent Sensor Management (ISM) reduce total cost of ownership compared to traditional sensors?

How does the pureO3 sensor support USP <1231> compliance in pharmaceutical water systems?