Silica measurement range
| 0-5000 ppb, limit of detection per ASTM Method D7126 |
Operating Range (Temp.)
| 5 °C – 60 °C 41 °F – 140 °F
|
Sample Flow rate
| 50 – 250 mL / min |
Process Connection
| Sample inlet: 6 mm or 1/4" OD tube SS compression fitting Drain hose: 19 x 25.4 mm (¾ x 1"), 2 m (6 ft) length included |
Protection Rating
| IP55 |
Ingress Protection
| IP55 (Enclosure) IP66 / NEMA4X (Electronics)
|
Silica Measurement Accuracy
| ± 5 % of reading or ± 1 ppb, whichever is greater |
Operator Interface
| Simultaneous display of silica concentration and measurement or auto-cal status TFT color touchscreen
|
Zero calibration
| Automatic, every measurement cycle |
Analog Output
| Powered 0 / 4 – 20 mA, 22 mA alarm, 500 ohm max load, not for use with externally powered circuit |
Analog Output Scaling
| Linear, bi-linear, logarithmic (1,2,3,4 decades), auto ranging |
Power Supply
| 100 – 240 VAC, 50 – 60 Hz, 100 W on power loss all settings are retained without batteries
|
Power Supply (AC/DC)
| 100 V – 240 V |
Flow Rate
| 50 mL/min – 250 mL/min |
Digital Input
| Remote start / stop of measurement cycle |
Measurement Parameter
| Silica |
Reagent Consumption
| Approx 500 mL each per 3 months with 20 min measurement cycle time |
Humidity
| 10 – 90 % non-condensing |
Communication
| PROFIBUS DP |
Grab Sample Measurement
| 500 mL capacity |
Sample Pressure
| 0.3 – 7 bar (5 – 100 psig) |
Analog Output Accuracy
| ± 0.05 mA |
Measurement Cycle Time
| 20 min typical Adjustable ≥ 10 min
|
Weight
| 18 kg (40 lbs) |
Sample Temperature
| 5 – 60 °C (41 – 140 °F) |
Model Family
| 2850Si |
EnclosureType
| Full |
Dimensions HWD
| Enclosure: 543 x 413 x 300 mm (21.4" x 16.3" x 11.8") |
Outputs
| 8 Powered 0 / 4 – 20 mA |
Relay Contact(s)
| 4-SPST Type Reed 250 VAC or DC, 0.5 A 4-SPST mechanical rated at 250 VAC, 3 A
|
Resolution
| 4 digits with decimal, autoranging 0.001 ppb in lowest range
|
Built-in sequencer
| Yes |
Ambient Operating Temperature
| 10 – 50 °C (50 – 122 °F) |
Ratings/Approvals
| CE, cULus |
Span calibration
| Automatic per schedule once per month, typical
|