Why Non-Aqueous Solvents Need a Different Measurement Approach

How the pH of Organic Solvents Was Measured

Assessing the Performance in Water-Miscible Organic Solvents

That said, in non-aqueous samples such as alcohol-based samples or ethanol fuel and fuel blends, ASTM D6423 recommends that the pH measurement is done at a timed endpoint of 30 seconds, upon immersion of the pH electrode in the sample.  

The following procedure is recommended to measure pH of samples dissolved in water-immiscible organic solvents:

icon

Take sample/ water-immiscible organic solvents and deionized water of controlled pH in a separating funnel and mix vigorously. The ratio of sample to water can be 1:1 or as per specified method. 

icon

Allow the mixture to stand until the aqueous and organic phases have completely separated.


icon

Collect the aqueous phase and measure its pH using standard aqueous measurement practices.

For this type of measurement, specialized sensors such as the InLab™ Science Pro-ISM and InLab™ Science Pro-DES can be used with a 3 mol/L KCl/water outer bridge electrolyte when measuring the extracted aqueous phase.

Why Direct Measurement Is Generally Not Recommended

When Direct Measurement Cannot Be Avoided

The Advantage of Choosing a Specialist Sensor for Non-Aqueous pH Measurements

Kindly note that the inner electrolyte is 3 mol/L KCl solution for this electrode and should not be changed, since the main reference system should remain Ag | AgCl | 3M KCl. 

To make the most of your pH measurement in organic solvents, follow these essential tips and best practices.

  • Ensure the use of fresh buffers for electrode calibration.
  • Make sure that the SafeLock (electrolyte filling hole) is open during pH calibration and measurement. For specialized sensors such as the InLab™ Science Pro-ISM and InLab™ Science Pro-DES, this guidance applies to both the inner and outer electrolyte chambers where applicable.
  • As a general guideline, additional care should be taken when measuring organic solvent systems containing more than 80% organic solvent. When measuring water-immiscible organic solvents, first rinse the sensor with a water-miscible solvent such as ethanol or isopropyl alcohol, and then rinse with water. This helps remove solvent residues more effectively and protects the sensor from contamination.
  • Thoroughly rinse the sensor after pH measurements in non-aqueous samples to prevent residue buildup and cross-contamination between measurements.

  • After exposure to organic solvents, allow the sensor to rehydrate by soaking it in an appropriate aqueous storage solution, such as InLab™ Storage Solution. Rehydration helps maintain the performance of the pH-sensitive glass membrane and supports reliable measurements.

  • Between measurements, or whenever the electrode is not in use, keep only the electrode tip (the pH-sensitive membrane) immersed in the storage solution. This helps recondition the glass membrane by enriching it with ions. Never store the sensor in ion-poor pure water.

  • Routine maintenance is essential for maximizing sensor performance and service life. Regularly inspect the sensor and replenish electrolyte solutions as required to ensure optimal performance.

Why is measuring pH in organic solvents challenging?

The traditional pH scale of 0 to 14 does not apply when measuring pH of organic solvents due to variations in how samples dissociate in different solvents. Organic solvents are often ion-deficient, leading to unstable measurements that require longer stabilization times. The aqueous outer electrolyte (3M KCl) is not suitable for use in an organic medium and must be replaced with an ethanolic solution of LiCl to obtain stable readings. Such samples require additional understanding to perform the pH measurement and sensor selection is key to obtain results with improved accuracy.

What is the InLab Science Pro-ISM pH sensor, and why is it suitable for measuring pH in organic solvents?

The InLab™ Science Pro-ISM pH sensor by METTLER TOLEDO is specifically designed for non-aqueous applications. The movable sleeve junction ensures easy and adequate outflow of reference electrolyte into the sample and is easy to clean in case of any blockage. This sensor has the added advantage of using two electrolytes, of which the outer bridge electrolyte can be changed from a conventional 3 mol/L KCl solution to 1 mol/L LiCl in an ethanol solution or any other electrolyte as needed. These factors contribute to obtaining stable readings and giving reliable pH results.

To learn more about the advantages of using this sensor, read this white paper.

Where is the pH of organic solvents measured and why is it important?

Several industries, including paints and pigments, inks, oils, petroleum and petrochemicals, pharmaceuticals, and biofuels, require routine pH measurements of samples that are dispensed or dissolved in organic solvents. Accurate pH measurement is crucial for maintaining product quality and compliance with industry standards.

How can I measure the pH of organic solvents?

pH measurements in organic solvents can be categorized into two sections: Organic solvents are classified as water-miscible and water-immiscible solvents. Therefore, pH measurements in non-aqueous samples can be categorized into two sections: 1. The pH measurement of water-miscible solvents, which is conducted by immersing the electrode in the solvent or its mixture (with water) 2. The pH measurement of water-immiscible organic solvents, which is conducted by performing a solvent extraction of the sample in the aqueous phase with deionized water, followed by the pH measurement

For further details, download our white paper on pH measurement in organic solvents.