Water-miscible organic solvents form a single phase with the aqueous system, but the presence of inorganic salts can lead to turbidity and precipitation in the mixture (especially if the concentration of the organic solvent is more than 80%).
A direct effect of this is noticed when a drop of 3 mol/L KCl solution, which is a conventional reference electrolyte, is added to the organic solvents. Solvents like ethanol, isopropanol, acetonitrile and DMSO, if present above 80%, precipitate out the KCl salt, because KCl is not as soluble in these solvents as it is in water.
In such a situation, the KCl electrolyte outflowing from the sleeve junction of the electrode precipitates and clogs the sensor junction. Hence, pH measurement in organic solvents using 3 mol/L KCl as the electrolyte is not advisable. 1 mol/L LiCl in ethanol solution can be used as the reference electrolyte in such cases. Indeed, LiCl has a much better solubility in ethanol than KCl (5.84 mol · kg−1 of solvent versus 0.0064 mol · kg−1 at 25 °C for KCl), while remaining perfectly soluble in water.
Additionally, the concentration is kept low at 1 mol/L to further improve solubility. The solvent medium for this electrolyte being ethanol helps miscibility in water-miscible organic solvents. The InLab Science Pro-ISM electrode enables the outer bridge electrolyte to 1 mol/L LiCl in ethanol for ensuring reliable pH measurement of non-aqueous samples in water miscible organic solvent.
Unlike KCl/water reference electrolyte, LiCl/ethanol fill solution is miscible in organic solvents in all proportions. This provides a stable junction potential during pH measurement, resulting in decreased response time and improved precision when compared with a conventional pH electrode.
The absolute pH values measured using KCl/water system would be different from LiCl/ethanol system due to the change in the overall reference system.