The water content of active pharmaceutical ingredients (APIs) is a key parameter in quality control, helping to ensure shelf life and proper function when used in medical treatment. KF titration is the method of choice for these water content determinations because it is selective for water.
To ensure accurate determination of water content, it is important that samples possess good solubility in KF solvents and release water completely. There are various methods to address poor solubility, including the introduction of solubilizers to the titration cell. However, this approach has limitations, including the potential for side reactions with the sample, and the risk of decreased cell conductivity if solubilizers are added beyond a certain percentage. This challenge can be overcome by introducing heat to the sample solution.
In this article, we will discuss a Karl Fischer (KF) titration method that can improve the solubility of sample solutions, leading to more precise and efficient analysis. Specifically, we will examine a direct addition technique that utilizes a thermostatable vessel. By adopting this technique, we can achieve better solubility, resulting in more complete water release and more accurate determination of water content in antibiotics.
Experimental
In order to increase the solubility of the Amoxicillin aliquot, the EVA V3 titrator and a special thermo vessel (thermostatable vessel, as shown in Figure 1) were used to perform the measurement. The temperature was set to 50 °C. After an initial heating period (~20 minutes), ~50 mg powdered sample was added directly into the titration cell with a weighing boat and the KF titration started. A mixing time of 120 seconds ensured dissolution, and the correct sample weight was obtained by back-weighing. This was repeated six times.
![]() |
Figure 1. EVA V3 Karl Fischer Titrator and a thermostatable vessel. |
Results and discussion
In order to dissolve the sample solution, a thermostatable vessel was utilized. However, applying heat has the potential to cause over-titration. To prevent this potential issue, the EVA titrator with the cautious titrant addition was used. By implementing these parameters, accurate titration results can be achieved.
The results of water content analysis for all six antibiotic aliquots using direct injection volumetric titration are presented in Table 1. It indicates a mean water content of 9.55 ± 0.25%.
According to the results displayed in Figure 2, the KF titration reactions of the six samples reached their endpoint within a range of 60–65 seconds. These results suggest that this technique enables fast and efficient analysis.
![]() |
Table 1. Results of water content for six Amoxycillin aliquots. |
![]() |
Figure 2. Titration curves of six Amoxycillin aliquots (current in µA vs Time in seconds). |
Conclusion
The METTLER TOLEDO EVA V3 volumetric Karl Fischer titrator is a highly effective solution for determining water content in challenging samples such as those that are insoluble. With its thermostable KF vessel and advanced features, this system delivers fast and reliable results with outstanding precision and accuracy.
Welcome everybody. Today I would like to show you a Calfisher titration application with our EVA titrators. It's water determination in antibiotics. Antibiotics are a relatively difficult sample because of their limited solubility in conventional Calfisher reagents solvents such as make the nose dry. That's why we use here heating in order to increase the solubility and to increase the water extraction. This thermostatic vessel is connected to a thermal state which will circulate the heating agent through this thermal vessel and normally one will use here 50 degree Celsius it takes about 20 minutes to heat up the Reagent, so this 20 minutes one has to wait after that one can add the antibiotics. One problem that can come up when heating coefficient reagent is the so-called over titration due to the increased mobility of the iodine and increased conductivity in the sample solution. So for this we have dedicated control parameters that you can select as cautious. These cautious parameters enable perfect titration water determination without over-titration. So now I would like to start the method and add the powder's antibiotics roughly 50 milligrams. So I start now here the method. And I will add here the antibiotic sample, roughly 50 milligrams. Now I will put back the weighing boat on the balance and the sample weight will automatically be transferred here onto the GUI and you can simply accept it now. And the titration will now start with a mix time of 60 seconds. This mix time is important in order to solubilize the antibiotics, but also to give enough time to liberate the water from the sample matrix. So this mix time is now going on. You see here a counter counting down from 60 seconds to zero. Seconds. You see also that measured value has dropped from 30 micro ampere which is the endpoint of this amperometric Koffischer titration to zero because all iodine has been consumed and now in the titration the titrator has again to reach the endpoint of 30 micro ampere and this will be done in a cautious way in order to Avoid any over titration. This is now going on. You can see here that the titrant consumption starts. We have a very high drift. This is because a lot of iodine is now added volumetrically here into the titration cell and the so far measured water content is already above 2.7 milligrams, reaching three milligrams and slowly the current is increasing here. You that more and more iodine is now again added into the cell, and as soon as we reach here the 30 microamperes, the titration will stop. We have as additional termination criteria the relative drift stop. Relative drift stop means that you add the drift to the start drift that you define in the termination criteria and then the titration will also stop. You always start from dry conditions when you do the titration. For this you define a certain allowed start drift. This may vary on the application type. So for some applications you may start with a higher start drift for other applications with a lower start drift. It is important that the drift remains constant. That's the most important part. Now we see here that Already 25 microgram microamperes have been reached and soon we reach here the end point of 30 microamps. The water content is above 5.6 milligrams and I expect now here the titration to complete soon. See here the drift coming down. And now we have here the final result, which should be between 9 and 11% for these antibiotics. So we are on the higher end, but it's perfectly reasonable result. This cautious titration achieved. And so we have here a nice solution using thermo vessel and dedicated titration parameters for this sample type. So thank you very much and see you soon.