DO Sensor InPro6970i/12/120

Optical DO Sensor for Brewery Industry.The InPro6970i trace optical oxygen sensor offers high operational availability, top performance and easy maintenance for the measurement of dissolved oxygen in brewery applications.

Withstands Challenging Applications

The sensor is designed for applications with a fast response time requirement. It is suitable for challenging ones such as filler line measurement.

High Operational Availability

The InPro 6970i offers outstanding measurement performance with a low detection limit of 0 ppb to 2000 ppb, minimum drift and a short response time.

Advanced Predictive Diagnostics

This DO sensor has Intelligent Sensor Management (ISM) technology, offering predictive diagnostics to reduce maintenance and simplify calibration.

Material No.: 52206393

Позвонить специалисту
Технические характеристики - DO Sensor InPro6970i/12/120
Измеряемый параметр
Кислород
Принцип измерения
Гашение флуоресценции
Время отклика (2)
Основная среда / область применения
Продукты питания и напитки
Цифровой (ISM) или аналоговый
Цифровой (ISM)
Материал металлических смачиваемых частей
Нержавеющая сталь 316L
Качество обработки поверхности смачиваемых частей
N5/Ra16 (Ra=0.4μm/16μin)
Диаметр датчика
12мм
Материал уплотнительного кольца
EPDM
Подключение к технологической линии
Pg 13.5
Длина
120 mm
Материал смачиваемой мембраны
Испольнение коннектора
прямой
Семейство моделей
Автоклавируемый/стерилизуемый
Да
Устойчивый к CIP
Да
Механическое отображение измерения
23 °F
250 °F
Рабочее давление, фунт/кв. дюйм
Интеграция с цифровыми схемами
Кабельный разъем
Сертификаты соответствия гигиеническим нормам / биологической совместимости
FDA/USP Class 6
Сертификаты и разрешения на применение
Диапазон измерения
0 - 2000 ppb насыщения
Рабочий диапазон (концентрация)
2 ppb
2 000 ppb
Сопротивление измеряемому давлению
0,2 bar
12 bar
Стерилизуемый, автоклавируемый
Предел чувствительности
2 ppb

Easy and Fast Operation

With the fast OptoCap replacement and the Plug & Measure ability, the InPro6970i optical oxygen sensor can be calibrated away from the process, and then quickly swapped into the process environment without the need to recalibrate. No polarization is required.

High Process Safety

The materials of construction of the InPro6970i trace DO sensor are FDA compliant and the easy-to-clean high-polished surface (N5 grade) also satisfies increasing regulatory requirements.

Enhanced Process Control

The sensor's measurement performance helps to reduce out of spec production. ISM helps to make optical oxygen measurement simple and more reliable and significantly reduces the risk of sensor failure during production.

ISM Redefines Maintenance

The built-in digital ISM technology provides the user important information about the sensor wear. Maintenance planning becomes easy, as accurate diagnostics predict when sensor maintenance will be required.

Документация

Технические характеристики

Manuals

Operating Instructions: Optical O₂ Simulator
METTLER TOLEDO optical O2 sensors are intended solely for inline measurement of the oxygen partial pressure, as described in this instruction manual.

Принадлежности

Средства калибровки и технического обслуживания

This is a USB iSense adaptor for connecting ISM RS485 sensors to iSense for pH, DO and CO₂ measurement and sensor calibration.

Material No.: 52300399

A universal device for connecting both 1-wire and RS485 ISM sensors to a computer running iSense software.

Material No.: 30130631

Расходные материалы

Программное обеспечение

Оформление запроса
DO Sensor InPro6970i/12/120
FDA
USP Class VI