In Situ FTIR Spectroscopy - Overview - METTLER TOLEDO

    In Situ FTIR Spectroscopy

    Understand Reaction Kinetics, Pathway and Mechanism with Real Time FTIR Spectroscopy

     
    Products and Specs
     
    For Use In
    Software
    Detector
    Optical Range (Base Unit)
    Purge Requirement
    Laboratory
    iC IR
    DTGS
    4000 – 650 cm-1
    No
    Material No.: 14474444
    For Use InLaboratory
    SoftwareiC IR
    DetectorDTGS
    Optical Range (Base Unit)4000 – 650 cm-1
    Purge RequirementNo
    Dimensions (WxHxD)137 mm x 116 mm x 241 mm
    Weight2.6 kg
    CommunicationEthernet TCP/IP
    Resolution4 cm -1 maximum
    Operating Range (Temp.)19 °C – 25 °C
    Probe Optical Window4000 – 650 cm-1
    Wetted MaterialsAlloy C-22;Diamond or Silicon;Gold
    Probe Pressure (Maximum)50bar
    Probe Temperature (Maximum)60°C
    Safety CertificationEMC Directive 2004/109/EC, Low Voltage Directive 2006/95/IEC
    Probe pH Range1 – 14 (Diamond);1 – 9 (Silicon)
    Laboratory
    iC IR
    LN2 NCT;SE MCT
    4000 – 650 cm-1
    No
    Material No.: 14000003
    For Use InLaboratory
    SoftwareiC IR
    DetectorLN2 NCT;SE MCT
    Optical Range (Base Unit)4000 – 650 cm-1
    Purge RequirementNo
    Dimensions (WxHxD)180 mm x 274 mm x 249 mm
    Weight20 lb (9 kg)
    CommunicationUSB 2.0
    Resolution4 cm-1 maximum
    Operating Range (Temp.)19 °C – 25 °C
    Probe Optical Window2500 – 650 cm-1 maximum (Fiber Probe)
    Wetted MaterialsAlloy C22;Diamond or Silicon;Gold
    Probe Pressure (Maximum)69bar (6.3 & 9.5mm Fiber Probe);107bar (Sentinel)
    Probe Temperature (Maximum)180°C (Fiber Probe);200°C (Sentinel)
    Safety CertificationEMC Directive 2004/109/EC, Low Voltage Directive 2006/95/IEC
    Probe pH Range1 – 14 (Diamond);1 – 9 (Silicon)
    Laboratory
    iC IR
    LN2 NCT
    4000 – 650 cm-1
    Yes
    Material No.: 14170003
    For Use InLaboratory
    SoftwareiC IR
    DetectorLN2 NCT
    Optical Range (Base Unit)4000 – 650 cm-1
    Purge RequirementYes
    Dimensions (WxHxD)216 mm x 381 mm x 279 mm
    Weight35 lb (16 kg)
    CommunicationUSB cable
    Resolution4 cm-1 maximum
    Operating Range (Temp.)19 °C – 25 °C
    Probe Optical Window2500 – 650 cm-1 maximum (Fiber Probe);4000 – 650 cm-1 (Sentinel)
    Wetted MaterialsAlloy C22;Diamond or Silicon;Gold
    Probe Pressure (Maximum)107bar (Sentinel);69bar (6.3 & 9.5mm Fiber Probe)
    Probe Temperature (Maximum)180°C (Fiber Probe);200°C (Sentinel)
    Safety CertificationEMC Directive 2004/109/EC, Low Voltage Directive 2006/95/IEC
    Probe pH Range1 – 14 (Diamond);1 – 9 (Silicon)
    Lab, Pilot Plant, or Production
    iC IR;iC Process
    DTGS;SE MCT
    4000 – 650 cm-1
    Yes
    Material No.: 14474487
    For Use InLab, Pilot Plant, or Production
    SoftwareiC IR;iC Process
    DetectorDTGS;SE MCT
    Optical Range (Base Unit)4000 – 650 cm-1
    Purge RequirementYes
    Dimensions (WxHxD)457 mm x 523 mm x 310 mm
    Weight83 lb (37 kg)
    CommunicationEthernet TCP/IP
    Resolution4 cm-1 maximum
    Probe Optical Window2500 – 650 cm-1 maximum (Fiber Probe);4000 – 650 cm-1 (Sentinel)
    Wetted MaterialsAlloy C22;Diamond or Silicon;Gold
    Probe Pressure (Maximum)107bar (Sentinel);69bar (6.3 & 9.5mm Fiber Probe)
    Probe Temperature (Maximum)180°C (Fiber Probe);200°C (Sentinel)
    Safety CertificationArea Classification – NON-HAZARDOUS, MET Mark E112462, UL61010-1 & CSA C22.2 No. 61010-1
    Probe pH Range1 – 14 (Diamond);1 – 9 (Silicon)
    Lab, Pilot Plant, or Production
    iC IR;iC Process
    DTGS;SE MCT
    4000 – 650 cm-1
    Yes
    Material No.: 14474485
    For Use InLab, Pilot Plant, or Production
    SoftwareiC IR;iC Process
    DetectorDTGS;SE MCT
    Optical Range (Base Unit)4000 – 650 cm-1
    Purge RequirementYes
    Dimensions (WxHxD)457 mm x 774 mm x 310 mm
    Weight100 lb (45 kg)
    CommunicationLC terminated Ethernet TCP/IP, duplex fiber optic
    Resolution4 cm-1 maximum
    Probe Optical Window2500 – 650 cm-1 maximum (Fiber Probe);4000 – 650 cm-1 (Sentinel)
    Wetted MaterialsAlloy C22;Diamond or Silicon;Gold
    Probe Pressure (Maximum)107bar (Sentinel);69bar (6.3 & 9.5mm Fiber Probe)
    Probe Temperature (Maximum)180°C (Fiber Probe);200°C (Sentinel)
    Safety CertificationArea Classification - HAZARDOUS, MET Mark E112462, UL61010-1 & CSA C22.2 No 61010-1 & NFPA 496, Class I, Division 1, Groups B;ATEX listing - TRAC12ATEX0001X, Equipment marking - Ex d op pr px [ia IIC] IIB+H2 T4 Gb;NFPA 496, Class I, Division 1, Groups B, T4
    Probe pH Range1 – 14 (Diamond);1 – 9 (Silicon)
    Comparison
     
     
     
     
     
     
     
    Thank you for visiting www.mt.com. We have tried to optimize your experience while on the site, but we noticed that you are using an older version of a web browser. We would like to let you know that some features on the site may not be available or may not work as nicely as they would on a newer browser version. If you would like to take full advantage of the site, please update your web browser to help improve your experience while browsing www.mt.com.