Measurement Parameter
| Oxygen |
Measurement Technology
| Amperometric/Polarographic |
Accuracy
| ≤ ± [1 % + 3 ppb] |
Response Time
| T98 = < 90 s @ 25°C (77°F)(air to N2) |
Primary Media/Application
| Food and Beverage |
IP Rating
| IP68 |
ISM (Digital) or Analog
| Analog |
Wetted Metallic Material
| 316L stainless steel |
Surface Roughnes of Wetted Material
| N5/Ra16 (Ra=0.4μm/16μin) |
Sensor Diameter
| 12mm |
O-Ring Material
| Silicone |
Process Connection
| Pg 13.5 |
Sensor Length
| 70 mm |
Wetted membrane material
| PTFE |
Connector Design
| Straight |
Model Family
| InPro 6900/6900i/6950i |
Autoclavable/Sterilizable
| Yes |
CIP resistant
| Yes |
Mechanical Temp. Range
| 23 °F – 249.8 °F (-5 °C – 121 °C) |
Operating Pressure (Bar)
| 0.2–6 Bar |
Operating Pressure (PSI)
| 0–87 PSI |
Design Pressure
| 174 psi (12 bar) |
Analog Integration
| nA Signal |
Cable Connector
| VP6 (IP68) |
Hygienic/Biocompatibility Approvals
| EHEDG/FDA/USP Class 7 |
Hazardous Area Approvals
| ATEX/IECEx: Ex ia IIC T6/T5/T4/T3 Ga/Gb Ex ia IIIC T69°C/T81°C/T109°C/T161°C Da/Db FM: IS / I, II, III / 1 / ABCDEFG/ T6 Ta=60 °C - 53 800 002 Entity
|
Certificates & Approvals
| Quality certificate, Material certificate 3.1, Surface finish certificate, ATEX/IECEx certificate, FDA/USP Class IV, EHEDG |
Sensor signal in air at 25 °C (77 °F)
| 250 to 500 nA |
Measurement Range
| 0-100% O2 saturation |
Operating Range
| 1 ppb - O2 sat. |
Sterilizable/Autoclavable
| Yes |
Residual signal in oxygen free media
| <0.03% of signal in ambient air |
Measured Pressure Resistance
| 0.2 bar – 9 bar |
Measuring Temp. Range
| 32 °F – 176 °F (0 °C – 80 °C) |
Detection limit
| 1 ppb |
Connection
| VP (IP68) |