The InPro6870i as an extension of INGOLD`s portfolio for optical oxygen sensors is designed for use in bioreactors where SIP is used for sanitization, whereas the InPro6880i can be autoclaved after the sensor head has been removed. Together with the new transmitter M400, ISM technology offers enhanced diagnostics to improve process safety drastically. Dynamic Lifetime Indicator (DLI) for replacement planning of the OptoCap and Adaptive Calibration Timer for highest accuracy during the process are only two of the new features to optimize the process. Additionally, the InPro6870i features a faster response time due to its temperature sensor now situated closer to the sensor tip, leading to a more accurate calibration and better process control. INGOLD`s optical oxygen sensors set a new standard for process safety.
If you still need OptCaps for InPro 6870i optical oxygen sensor, Please visit here.
Increased Yield and Product Quality
The angled tip and hydrophilic surface prevent bubble formation. This eliminates noise for improved production efficiency and process control.
For Digital & Analog Bioreactors
This anti-bubble DO sensor has digital (MODBUS, RS-485) and analog (mA) output signals for compatibility with benchtop bioreactors and transmitters.
Sterilizable and Autoclavable
The InPro 6860i/12/120/mA/HD provides fast, precise oxygen measurement, even after repeated autoclaving and sterilization cycles up to 140° C.